Compur 4120 Switch positions of the sensor head pcb, Slide switch S1, Toggle switch S3

Page 15

Compur Statox 4120

Slide switch S1:

 

S1

 

ON

 

 

OFF

 

 

 

 

 

 

 

1

 

Relay active during alarm A2

 

 

Relay inactive at alarm A2

 

 

2

 

Relay active during alarm A1

 

 

Relay inactive at alarm A1

 

Toggle switch S3:

S3

ON

OFF

1

Detection mode

-

2

Rest Alarm by resetbutton R

auto - reset

3

Detection mode

-

4

Detection mode

-

5

Measured value accepted if logic

-

 

signal is 0V

 

6

Detection mode

-

7

Detection mode

Modem - and analog signal

 

 

adjustment for service only!

8

Detection mode

Delete EPROM for service only!

Ex works settings

3.4 Switch positions of the sensor head pcb

DIP switch S2 (see pic. 9) programs the following parameters:

S2

ON

OFF

1

-

Detection mode

2

Measured value in ppm

Measured value in TLV

3

No self test after “power on“

Self test after “power on“

4

Cal. Factor =1

Factor information from control

 

 

module

5

No self test

Self test in a 24 h interval

6

-

-

7

Detection mode

-

8

Detection mode

Modem “On” for service

Ex works setting

15

Image 15
Contents Compur Statox Compur Statox Compur Statox 4120 operation Installation and electrical connectionsTrouble shooting Maintenance and calibrationStatox 4120 system Purpose of use and description Important noticeStatox 4120 sensor head Statox 4120 rack Operation principleAreas Connections ConstructionLED SF 8 „System Fail Statox 4120 control moduleBargraph display LED SR 7 „System readyDiagnostic box Installation and electrical connections Statox 4120 sensor headMounting Connecting the signal cable Sensor head connection procedure Example of cable specificationsStatox 4120 rack Installation instructionsStatox 4120 control module Switch positions of the sensor head pcb Slide switch S1Toggle switch S3 Switch position for H2S Switch position for HCNSwitch position for COCl2 Switch position for Cl2 Switch position for COSwitch position for HCl Switch position for ClO2 Start up Alarm thresholdsDetection mode Compur Statox 4120 operationSystem failure Trouble shootingSF flashing Manually triggered self test of the systemDiagnostic box error messages Using the diagnostic boxOperation and error codes of the diagnostic box Diagnostic box queries Button DisplayCompur Statox Inspection Statox 4120 Sensor head Maintenance and calibrationSensor head calibration Filter cartridge replacementGenerator cell replacement Using span gasProcedure This is the last opportunity to abort the procedureUse ex works calibration Maintenace request option Key Mode Relay PCS Relay WB Analog signal Reading Error C3 Too low sensitivity during self testMounting the maintenance request pcb Service mode control moduleBridge Analog 20 mA 40-20mA Analog Terminal Solder straps Option analog output inhibitionSystem requirements Sensor head Technical data General Gas Measuring rangeDiagnostic box Control moduleRack Accessories and spare parts Art. NrArt.Nr Sensors & Generator Cells 558435