Compur Statox 4120
5.1 Using the diagnostic box
Switch the diagnostic box on. Then connect it to the opto- coupler receptacle on sensor head bottom. Keep the LED`s and the phototransistors of the opto- coupler clean for proper data transmission.
5.1.1 Operation and error codes of the diagnostic box
Note: If a combination of push button operations is requested ( f. i. ST + ENT), the buttons must be operated in a short time distance.
Example: ST + ENT initiates a self test. The display says
Diagnostic box queries:
|
| Button | Display |
| ||
|
| M | Measured value in ppm |
| ||
|
| M + ENT | WAIT; after a while reading of actual measured value in ppm |
| ||
|
| RA | Sensor head measuring range |
| ||
|
| RA + ENT | Gas |
| ||
|
| ST | Status passed (System OK) or error message (see 5.1.2) |
| ||
|
| ST + ENT | (initiates self test), TEST Run |
| ||
|
| FKT | Calibration factor |
| ||
|
| ZERO + ENT * | ZERO RUN (start zero adjustment) |
| ||
|
| * only available with systems with zero adjustment | ||||
Diagnostic box error messages: | ||||||
|
|
|
|
| ||
| ERR INPU |
|
| Faulty or too slow input. Repeat. | ||
|
|
|
|
| ||
| ERR KONZ |
|
| Calibration gas concentration out of range. | ||
|
|
|
|
| ||
| ERR FACT |
|
| Calibration factor is out of range. | ||
|
|
|
|
| ||
| DS.W.POS |
|
| Measuring range or gas not correctly programmed. | ||
|
|
|
|
| ||
| WARN TIM |
|
| Time error during data transmission. | ||
|
|
|
|
| ||
| WARN DTE |
|
| Faulty data transmission. | ||
|
|
|
|
| ||
| WARN CHE |
|
| No self test due to too low temperature. | ||
|
|
|
|
| ||
| TEMP.OU |
|
| Temperature out of the admissible range. | ||
|
|
|
|
| ||
| TEMP.IN |
|
| Temperature is back in the admissible range. | ||
|
|
|
|
|
| |
| CH.BATT |
|
| Charge calibration box battery. | ||
| (audible alarm) |
|
| |||
|
|
|
|
| ||
| NO ZERO |
|
| Zero adjustment not possible. Wrong EPROM in sensor | ||
|
|
| head and control module. | |||
|
|
|
|
| ||
| PUMP RUN |
|
| Sensor is flushed with air because the gas concentration is | ||
|
|
| too high. | |||
|
|
|
|
| ||
| CAL END |
|
| Calibration routine has been cancelled with CE button. | ||
|
|
|
|
| ||
| NO CALIB |
|
| Calibration not possible within the time limit. | ||
|
|
|
|
|
|
|
20