Compur 4120 Using the diagnostic box, Operation and error codes of the diagnostic box

Page 20

Compur Statox 4120

5.1 Using the diagnostic box

Switch the diagnostic box on. Then connect it to the opto- coupler receptacle on sensor head bottom. Keep the LED`s and the phototransistors of the opto- coupler clean for proper data transmission.

5.1.1 Operation and error codes of the diagnostic box

Note: If a combination of push button operations is requested ( f. i. ST + ENT), the buttons must be operated in a short time distance.

Example: ST + ENT initiates a self test. The display says „WAIT-TEST RUN“ until the self test is finished.

Diagnostic box queries:

 

 

Button

Display

 

 

 

M

Measured value in ppm

 

 

 

M + ENT

WAIT; after a while reading of actual measured value in ppm

 

 

 

RA

Sensor head measuring range

 

 

 

RA + ENT

Gas

 

 

 

ST

Status passed (System OK) or error message (see 5.1.2)

 

 

 

ST + ENT

(initiates self test), TEST Run

 

 

 

FKT

Calibration factor

 

 

 

ZERO + ENT *

ZERO RUN (start zero adjustment)

 

 

 

* only available with systems with zero adjustment

Diagnostic box error messages:

 

 

 

 

 

 

ERR INPU

 

 

Faulty or too slow input. Repeat.

 

 

 

 

 

 

ERR KONZ

 

 

Calibration gas concentration out of range.

 

 

 

 

 

 

ERR FACT

 

 

Calibration factor is out of range.

 

 

 

 

 

 

DS.W.POS

 

 

Measuring range or gas not correctly programmed.

 

 

 

 

 

 

WARN TIM

 

 

Time error during data transmission.

 

 

 

 

 

 

WARN DTE

 

 

Faulty data transmission.

 

 

 

 

 

 

WARN CHE

 

 

No self test due to too low temperature.

 

 

 

 

 

 

TEMP.OU

 

 

Temperature out of the admissible range.

 

 

 

 

 

 

TEMP.IN

 

 

Temperature is back in the admissible range.

 

 

 

 

 

 

 

CH.BATT

 

 

Charge calibration box battery.

 

(audible alarm)

 

 

 

 

 

 

 

 

NO ZERO

 

 

Zero adjustment not possible. Wrong EPROM in sensor

 

 

 

head and control module.

 

 

 

 

 

 

PUMP RUN

 

 

Sensor is flushed with air because the gas concentration is

 

 

 

too high.

 

 

 

 

 

 

CAL END

 

 

Calibration routine has been cancelled with CE button.

 

 

 

 

 

 

NO CALIB

 

 

Calibration not possible within the time limit.

 

 

 

 

 

 

 

20

Image 20
Contents Compur Statox Compur Statox Installation and electrical connections Trouble shootingMaintenance and calibration Compur Statox 4120 operationImportant notice Statox 4120 system Purpose of use and descriptionStatox 4120 sensor head Areas Statox 4120 rackOperation principle Construction ConnectionsStatox 4120 control module Bargraph displayLED SR 7 „System ready LED SF 8 „System FailDiagnostic box Mounting Installation and electrical connectionsStatox 4120 sensor head Connecting the signal cable Example of cable specifications Sensor head connection procedureInstallation instructions Statox 4120 rackStatox 4120 control module Toggle switch S3 Switch positions of the sensor head pcbSlide switch S1 Switch position for COCl2 Switch position for H2SSwitch position for HCN Switch position for HCl Switch position for ClO2 Switch position for Cl2Switch position for CO Alarm thresholds Detection modeCompur Statox 4120 operation Start upTrouble shooting SF flashingManually triggered self test of the system System failureUsing the diagnostic box Operation and error codes of the diagnostic boxDiagnostic box queries Button Display Diagnostic box error messagesCompur Statox Maintenance and calibration Inspection Statox 4120 Sensor headFilter cartridge replacement Generator cell replacementUsing span gas Sensor head calibrationUse ex works calibration ProcedureThis is the last opportunity to abort the procedure Maintenace request option Error C3 Too low sensitivity during self test Mounting the maintenance request pcbService mode control module Key Mode Relay PCS Relay WB Analog signal ReadingBridge Analog 20 mA 40-20mA Option analog output inhibition Analog Terminal Solder strapsSystem requirements Technical data General Gas Measuring range Sensor headRack Diagnostic boxControl module Art.Nr Accessories and spare partsArt. Nr Sensors & Generator Cells 558435