Compur operation manual Statox 4120 control module, Bargraph display, LED SR 7 „System ready

Page 8

Compur Statox 4120

2.3 Statox 4120 control module

Each Statox 4120 Control module has its own power supply. The intrinsically safe sensor head supply circuit is supplied by a separate secondary winding.

 

pic.4: Statox 4120 Rack

1

19“ Rack

9

LED A1

2

Statox 4120 control module

10

LED A2

3

Button A2

11

Potentiometer for A2

4

Button A1

12

Potentiometer A1

5

Button T

13

Bar graph display

6

Button R

14

Blind panel

7

LED SR

15

Rail

8

LED SF

16

ppm Scale

Bargraph display (13)

-Actual measured concentration value.

-Flashing when measuring range is exceeded and pump is on.

LED SR (7) „System ready“

-On if system is working properly.

-Flashing after power up until system is communicating fault - free.

-Flashing while diagnostic box is connected.

LED SF (8) „System Fail“

-On if a system failure has occurred. (Self test not passed, communication error, cable interrupted).

8

Image 8
Contents Compur Statox Compur Statox Installation and electrical connections Trouble shootingMaintenance and calibration Compur Statox 4120 operationImportant notice Statox 4120 system Purpose of use and descriptionStatox 4120 sensor head Areas Statox 4120 rackOperation principle Construction ConnectionsStatox 4120 control module Bargraph displayLED SR 7 „System ready LED SF 8 „System FailDiagnostic box Mounting Installation and electrical connectionsStatox 4120 sensor head Connecting the signal cable Example of cable specifications Sensor head connection procedureInstallation instructions Statox 4120 rackStatox 4120 control module Toggle switch S3 Switch positions of the sensor head pcbSlide switch S1 Switch position for COCl2 Switch position for H2SSwitch position for HCN Switch position for HCl Switch position for ClO2 Switch position for Cl2Switch position for CO Alarm thresholds Detection modeCompur Statox 4120 operation Start upTrouble shooting SF flashingManually triggered self test of the system System failureUsing the diagnostic box Operation and error codes of the diagnostic boxDiagnostic box queries Button Display Diagnostic box error messagesCompur Statox Maintenance and calibration Inspection Statox 4120 Sensor headFilter cartridge replacement Generator cell replacementUsing span gas Sensor head calibrationUse ex works calibration ProcedureThis is the last opportunity to abort the procedure Maintenace request option Error C3 Too low sensitivity during self test Mounting the maintenance request pcbService mode control module Key Mode Relay PCS Relay WB Analog signal ReadingBridge Analog 20 mA 40-20mA Option analog output inhibition Analog Terminal Solder strapsSystem requirements Technical data General Gas Measuring range Sensor headRack Diagnostic boxControl module Art.Nr Accessories and spare partsArt. Nr Sensors & Generator Cells 558435