Compur 4120 operation manual Option zero adjustment, System requirements

Page 29

Compur Statox 4120

9.The option zero adjustment

This option is available for COCl2 sensor heads with measuring range 0.3 ppm only. The zero adjustment of the sensor is started via diagnostic box.

System requirements:

Sensor head EPROM index 18c, control module EPROM index 03a, diagnostic box EPROM index 02.

A zero adjustment must only be done in clean air. It is started with the buttons ZERO + ENT. It should be done before starting a calibration.

The set zero value is not visible.

If no zero adjustment has been done ever, the default value is 0.

After replacing a control module, a new zero setting has to be carried out. The zero value is stored in the control module and is transmitted to the sensor head.

An error message C1 caused by a too high sensor zero current will only be cancelled after a successful self test.

29

Image 29
Contents Compur Statox Compur Statox Trouble shooting Installation and electrical connectionsMaintenance and calibration Compur Statox 4120 operationStatox 4120 system Purpose of use and description Important noticeStatox 4120 sensor head Areas Statox 4120 rackOperation principle Connections ConstructionBargraph display Statox 4120 control moduleLED SR 7 „System ready LED SF 8 „System FailDiagnostic box Mounting Installation and electrical connectionsStatox 4120 sensor head Connecting the signal cable Sensor head connection procedure Example of cable specificationsStatox 4120 rack Installation instructionsStatox 4120 control module Toggle switch S3 Switch positions of the sensor head pcbSlide switch S1 Switch position for COCl2 Switch position for H2SSwitch position for HCN Switch position for HCl Switch position for ClO2 Switch position for Cl2Switch position for CO Detection mode Alarm thresholdsCompur Statox 4120 operation Start upSF flashing Trouble shootingManually triggered self test of the system System failureOperation and error codes of the diagnostic box Using the diagnostic boxDiagnostic box queries Button Display Diagnostic box error messagesCompur Statox Inspection Statox 4120 Sensor head Maintenance and calibrationGenerator cell replacement Filter cartridge replacementUsing span gas Sensor head calibrationUse ex works calibration ProcedureThis is the last opportunity to abort the procedure Maintenace request option Mounting the maintenance request pcb Error C3 Too low sensitivity during self testService mode control module Key Mode Relay PCS Relay WB Analog signal ReadingBridge Analog 20 mA 40-20mA Analog Terminal Solder straps Option analog output inhibitionSystem requirements Sensor head Technical data General Gas Measuring rangeRack Diagnostic boxControl module Art.Nr Accessories and spare partsArt. Nr Sensors & Generator Cells 558435