Compur operation manual Statox 4120 rack, Operation principle, Areas

Page 6

Compur Statox 4120

pic. 2: Statox 4120 sensor head for Cl2 und ClO2

Operation principle

Ambient air enters the sensor (2) by diffusion through an opening at the bottom. The sensor creates a current - signal proportional to the gas concentration. This signal is amplified, temperature compensated and linearised. As a digital signal it is transmitted to the control module.

To protect the sensor from being poisoned by a high gas concentration, the Statox features a built – in air purge. This is activated when the gas concentration exceeds 95 % of the measuring range and stops when it has dropped below 80 %.

The battery pack (7) buffers the energy demand of the built – in pump (5). It is on permanent trickle charge. The pump is activated in the protection mode and during the self test to purge the sensor and provide test gas from the gas generator (4) to the sensor.

2.2 Statox 4120 rack

The Statox 4120 rack and the Statox 4120 control module provide the power supply for the sensor head. They are certified explosion proof [EEx ib] IIC. The rack must not be installed in certified

areas!

6

Image 6
Contents Compur Statox Compur Statox Maintenance and calibration Installation and electrical connectionsTrouble shooting Compur Statox 4120 operationImportant notice Statox 4120 system Purpose of use and descriptionStatox 4120 sensor head Statox 4120 rack Operation principleAreas Construction ConnectionsLED SR 7 „System ready Statox 4120 control moduleBargraph display LED SF 8 „System FailDiagnostic box Installation and electrical connections Statox 4120 sensor headMounting Connecting the signal cable Example of cable specifications Sensor head connection procedureInstallation instructions Statox 4120 rackStatox 4120 control module Switch positions of the sensor head pcb Slide switch S1Toggle switch S3 Switch position for H2S Switch position for HCNSwitch position for COCl2 Switch position for Cl2 Switch position for COSwitch position for HCl Switch position for ClO2 Compur Statox 4120 operation Alarm thresholdsDetection mode Start upManually triggered self test of the system Trouble shootingSF flashing System failureDiagnostic box queries Button Display Using the diagnostic boxOperation and error codes of the diagnostic box Diagnostic box error messagesCompur Statox Maintenance and calibration Inspection Statox 4120 Sensor headUsing span gas Filter cartridge replacementGenerator cell replacement Sensor head calibrationProcedure This is the last opportunity to abort the procedureUse ex works calibration Maintenace request option Service mode control module Error C3 Too low sensitivity during self testMounting the maintenance request pcb Key Mode Relay PCS Relay WB Analog signal ReadingBridge Analog 20 mA 40-20mA Option analog output inhibition Analog Terminal Solder strapsSystem requirements Technical data General Gas Measuring range Sensor headDiagnostic box Control moduleRack Accessories and spare parts Art. NrArt.Nr Sensors & Generator Cells 558435