Compur 4120 Switch position for H2S, Switch position for HCN, Switch position for COCl2

Page 16

Compur Statox 4120

DIP switch S1 (see pic. 8) programs gas and measuring range. If you change your Statox to another gas or measuring range, a hardware modification may be required and the electronics must be re - adjusted. This must be done by authorised personnel.

Table 1: Switch position for H2S

Measuring range:

0-30 ppm

0-50 ppm

0-100 ppm

S1/1

ON

OFF

OFF

S1/2

ON

ON

ON

S1/3

ON

ON

ON

S1/4

ON

ON

OFF

S1/5

ON

ON

OFF

S1/6

ON

OFF

ON

S1/7

ON

ON

OFF

S1/8

ON

ON

ON

E-PROM Index sensor head

12, 13, 14

12, 13, 14

12, 13, 14

E-PROM Index control module

≥ 02

≥ 02

≥ 02

Table 2: Switch position for HCN

Measuring range:

0-15 ppm

0-30 ppm

0-50 ppm

0-100 ppm

S1/1

ON

ON

OFF

OFF

S1/2

ON

ON

ON

ON

S1/3

ON

ON

ON

ON

S1/4

OFF

OFF

OFF

ON

S1/5

OFF

ON

ON

ON

S1/6

OFF

ON

OFF

ON

S1/7

OFF

ON

ON

ON

S1/8

ON

ON

ON

ON

E-PROM Index sensor head

13

12, 13, 14

12, 13, 14

15

E-PROM Index control module

≥ 02

≥ 02

≥ 02

≥ 02

Table 3: Switch position for COCl2

 

Measuring range:

 

 

0-0,1 ppm

 

 

0-0,3 ppm

 

 

0-0,3 ppm

 

 

0-0,5 ppm

 

 

0-1,5 ppm

 

 

0-15 ppm

 

 

0-100 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S1/1

 

 

OFF

 

 

ON

 

 

ON

 

 

OFF

 

 

OFF

 

 

OFF

 

 

OFF

 

 

S1/2

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

S1/3

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

S1/4

 

 

OFF

 

 

ON

 

 

ON

 

 

ON

 

 

OFF

 

 

ON

 

 

ON

 

 

S1/5

 

 

ON

 

 

OFF

 

 

ON

 

 

OFF

 

 

ON

 

 

OFF

 

 

OFF

 

 

S1/6

 

 

ON

 

 

ON

 

 

ON

 

 

OFF

 

 

OFF

 

 

OFF

 

 

ON

 

 

S1/7

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

OFF

 

 

OFF

 

 

ON

 

 

S1/8

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

ON

 

 

E-PROM Index

 

20

 

12, 13, 14

 

18

 

12, 13, 14

 

12, 13, 14

 

12, 13, 14

 

17

 

 

sensor head

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

E-PROM Index

 

04

 

 

02

 

 

02

 

 

02

 

 

02

 

 

02

 

 

02

 

 

control module

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

16

Image 16
Contents Compur Statox Compur Statox Installation and electrical connections Trouble shootingMaintenance and calibration Compur Statox 4120 operationImportant notice Statox 4120 system Purpose of use and descriptionStatox 4120 sensor head Operation principle Statox 4120 rackAreas Construction ConnectionsStatox 4120 control module Bargraph displayLED SR 7 „System ready LED SF 8 „System FailDiagnostic box Statox 4120 sensor head Installation and electrical connectionsMounting Connecting the signal cable Example of cable specifications Sensor head connection procedureInstallation instructions Statox 4120 rackStatox 4120 control module Slide switch S1 Switch positions of the sensor head pcbToggle switch S3 Switch position for HCN Switch position for H2SSwitch position for COCl2 Switch position for CO Switch position for Cl2Switch position for HCl Switch position for ClO2 Alarm thresholds Detection modeCompur Statox 4120 operation Start upTrouble shooting SF flashingManually triggered self test of the system System failureUsing the diagnostic box Operation and error codes of the diagnostic boxDiagnostic box queries Button Display Diagnostic box error messagesCompur Statox Maintenance and calibration Inspection Statox 4120 Sensor headFilter cartridge replacement Generator cell replacementUsing span gas Sensor head calibrationThis is the last opportunity to abort the procedure ProcedureUse ex works calibration Maintenace request option Error C3 Too low sensitivity during self test Mounting the maintenance request pcbService mode control module Key Mode Relay PCS Relay WB Analog signal ReadingBridge Analog 20 mA 40-20mA Option analog output inhibition Analog Terminal Solder strapsSystem requirements Technical data General Gas Measuring range Sensor headControl module Diagnostic boxRack Art. Nr Accessories and spare partsArt.Nr Sensors & Generator Cells 558435