Miele G 7824, G 7823 operating instructions Operation

Page 42

Operation

Master switch

The master switch with "emergency cut-off function" is set to I-ON.

Switching on

￿Open the stopcocks (if turned off).

￿Press the I-0button.

In operating levels A, B and D the most recently selected PROGRAMME NAME appears in the display and in operating level C AUTOMATIC MOBILE UNIT RECOGNITION is displayed.

Door lock

The door can only be opened if:

the electricity supply to the machine is switched on,

the red (emergency cut out) master switch is in the I-ONposition,

the I-0On-Off button is pressed in, and

a cleaning or disinfection programme is not in progress.

To open the door

￿Press the ￿ door button, and holding onto the door grip open the door.

To open the door on the Clean side (through-feed model), the following conditions must also be met:

DESIN (disinfection) programmes have been correctly completed,

or

the door locking function is active (see "Machine function/Door Interlock" in the Programming manual).

￿Do not touch the heating elements under the wide area filter during or directly after the end of a programme. You could burn yourself. They remain hot for some time after the programme has finished.

If the boiler has been programmed to BOILER READY, be particularly careful of hot water and steam when opening the door. Danger of burning or scalding. The inlet to the wash cabinet is located, as viewed from the unclean side, underneath the basket runner on the left hand side.

To close the door

Lift the door upwards and push until it clicks shut.

42

Image 42
Contents Washer-disinfector G 7823 / G Contents Contents Correct application Page Page Using accessories Symbols on the appliance Disposing of your old machineIntended Use Examples of application areasUnclean side Guide to the machineGuide to the machine Clean side G 7824 only Controls DisplayCursor button right Moves the cursor to the right Door switch Start buttonService interface Programme selector Minus buttonSetting the water softener Water softener softener OptionalReactivation display Reactivating the water softener Water softener softener Optional Run the reactivation programme ThenAutomatic mobile unit recognition Optional Mobile unit coding Setting mobile unit coding Mobile unit coding stripSet the programme place coding Areas of application Protein testAt the end of each programme Loading the machine Preparing the loadUnloading instruments Surgical instruments OP Ophthalmology Anaesthetic instruments AN Baby bottles BC Loading in an EBottom level with 3 x E Frame Operating theatre shoes OS Transfer trolley for loading and unloading the machine Areas of application General notes Problem How to resolve it Chemical processes and technologyChemical agents Problem How to resolve it Chemical processes and technology Soiling Problem How to resolve it Chemical processes and technology Oxivario kit Dispensing liquid chemical agentsAdditional optional DOS modules Orthovario kitPreparing the DOS dispensers Fill the containers with the relevant agentDispensing liquid chemical agents External Dispensing systems Dispensing system maintenanceVenting the dispensing system Every 12-18 monthsOperation Operating levels Selecting or changing your own codeStarting a programme Fixed programmeFree programme selection Programme sequence AWK Automatic mobile unit recognitionSwitching off Cancelling a programmeOperating levels B or D Interrupting a programme Data transfer Maintenance Periodic checksProcess validation Routine checksCleaning the filters in the wash cabinet Cleaning the coarse filterCleaning the coarse, flat and micro-fine filters Cleaning the spray arms Cleaning the wash cabinet Cleaning the control panelCleaning the front of the machine Cleaning the door sealsMobile units, modules and inserts Printer optional Renewing the paper rollRenewing the colour cassette After sales service Converting the type of heating Electrical connection Plumbing Technical data Optional extras Optional extrasSpecial processes Oxivario and Orthovario Description of the machineAreas of application Connecting the H2O2 solution container Disposal of your old machine Page 0366