Special processes - OXIVARIO and ORTHOVARIO
Warning and Safety instructions
These warning and safety instructions are in addition to those given at the beginning of this booklet.
H2O2solution must only be used in the special containers and adapters provided by the Ecolab and Dr. Weigert companies.
Important: observe the manufacturer's safety instructions as given on their safety data sheet.
Be especially careful when handling H2O2 solution. It is an irritant chemical. Observe all current safety instructions pertaining to its use.
Wear protective gloves and goggles.
Empty containers must be disposed of in accordance with the manufacturer's instructions.
H2O2 solution must not be mixed
with other chemicals. This could cause a serious chemical reaction, e. g. the release of harmful vapours or gases.
Only use tenside cleaning agents supplied by Ecolab or Dr. Weigert.
Connecting the H2O2 solution container
The connection hose for the H2O2 solution container is identified by a black label. The connection hose is supplied without a safety cap as the container for hydrogen peroxide will vary depending on supplier.
Connect the supplier's safety cap to the connection hose (black).
Then connect this to the H2O2 solution container.
Then start the
Unlike with other chemical agents the H2O2 solution container must be fully emptied before it is exchanged for a new one.
Do not replace the H2O2 solution container with a new one until the FILL
DOS 2 CONTAINER message is
displayed. After connecting the new container you must run the programme.
If the CHECK DISPENSING SYSTEM 2
message appears you should check the container as well as the dispensing system. The programme will stop automatically.
65