| INSTALLATION 3 |
| SITRANS FX300 | |
|
|
|
3.4.1 Measurement of liquids
Prohibited installation
Figure 3-2: Upper pipe bend
CAUTION!
Prohibited: Installing the device in an upper pipe bend 1, because there is a risk of gas bubbles 2 forming. Gas bubbles can lead to pressure surges and inaccurate measurement.
Figure 3-3: Downpipe and outlet
CAUTION!
Prohibited: Installing the device in a downstream pipe 3 or upstream pipe of a outlet 4. There is the risk of partially filled pipes.
06/2009 - SFIDK.PS.050.F2.02 A5E02100423 | www.siemens.com/flow | 19 |