6 OPERATION |
|
|
SITRANS FX300 |
| |
|
|
|
6.5.7 Menu item Setup (firmware version - steam)
Level | Designation | Selection / entry | Explanation | ||
|
|
|
|
|
|
3.1 |
|
|
|
| Set display |
|
|
|
|
|
|
3.1.1 | Error Msg. |
|
|
| Display error |
| → |
|
|
|
|
| Yes | ↑... |
| Shows error messages in measuring mode in plain text | |
|
|
| |||
|
|
|
|
| alternating with the measured values |
|
|
|
|
| |
|
| No | ↑...^ | Flashing cursor in the top left corner of the display | |
|
|
|
|
| indicates errors. |
3.1.2 | Cycle.Disp. |
|
|
| Display measured values |
| → |
|
|
|
|
| Yes | ↑... |
| Cyclically display measured values in measuring mode | |
|
|
| |||
|
|
|
|
| (6 s intervals) |
|
|
|
|
| |
|
| No | ↑...^ | Do not display measured values cyclically | |
|
|
|
|
|
|
3.1.3 | Code 1 |
|
|
| Set security query 1 (menu) |
| → |
|
|
|
|
| Yes | ↑... |
| Activate access code: | |
|
|
| |||
|
|
|
|
| →→→↑↑↑^^^ |
|
|
|
|
| |
|
| No | ↑...^ | Deactivate access code | |
|
|
|
|
|
|
|
|
|
|
|
|
3.2 |
|
|
|
| I/O settings |
|
|
|
|
|
|
3.2.1 |
|
|
| Current output | |
| → |
|
|
|
|
| Current output | ||||
|
| ||||
|
|
|
|
| |
|
| ↑... | Current output | ||
|
|
|
| ||
|
| Current output | |||
|
|
|
|
|
|
3.2.2 | Variable I |
|
|
| Set output variable for current output |
| → |
|
|
|
|
| Maes.Inst ↑... | Flow rate | |||
| (available if Meter Type = | ||||
| Gross Heat, see menu | Power ↑...^ | Power | ||
| item 5.4.1, service menu) | ||||
|
|
|
|
| |
|
|
|
|
|
|
3.2.3 | Power Unit |
|
|
| Set unit for heat output |
| → |
|
|
|
|
| kJ/h |
|
| Select unit for measuring heat output | |
| (available if Meter Type = | ↑...^ | |||
| Unit |
| |||
| Gross Heat, see menu |
| |||
| item 5.4.1, service menu) | Disp. on / | Display measured value/ | ||
|
| Disp. off | ↑...^ | Do not display measured value | |
|
|
|
|
|
|
3.2.4 | FS power |
|
|
| End value for power measuring range |
| → |
|
|
|
|
| 7000000.00 | Power value at 20 mA current output | |||
| (available if Variable I = | ||||
| kJ/h ^ |
|
| ||
| Power, see menu item |
|
| ||
| 3.2.2) |
|
|
|
|
|
|
|
|
|
|
3.2.5 | Function P |
|
|
| Function P |
| → |
|
|
|
|
| Yes | ↑... |
| Activate pulse output | |
|
|
| |||
|
|
|
|
| |
|
| No | ↑...^ | Deactivate pulse output | |
|
|
|
|
|
|
3.2.6 | Variable P |
|
|
| Set output variable for the pulse output |
| → |
|
|
|
|
| Total Flow ↑... | Total flow | |||
| (available if Meter Type = | ||||
| Gross Heat, see menu | Energy ↑...^ | Energy | ||
| item 5.4.1, service menu) | ||||
|
|
|
|
| |
|
|
|
|
|
|
3.2.7 | Totalizer |
|
|
| Totalizer |
| → |
|
|
|
|
| Tot. on | ↑... | Start totalizer | ||
|
| ||||
|
|
|
|
| |
|
| Tot. off | ↑...^ | Stop totalizer (value is held) | |
|
|
|
|
|
|
3.2.8 | Tot. Conf. |
|
|
| Configure totalizer |
| → |
|
|
|
|
|
|
|
|
|
|
54 | www.siemens.com/flow | 06/2009 - SFIDK.PS.050.F2.02 A5E02100423 |