Siemens FX300 manual Electrical Connections, Following procedures are to be performed

Page 32

4 ELECTRICAL CONNECTIONS

 

 

 

 

SITRANS FX300

 

4.2 Connecting the signal converter

 

 

 

 

Figure 4-1: Signal converter housing with housing cover

1Housing cover of the electrical terminal compartment

2Electrical connection terminals with the housing cover open

3Terminal A current output -

4Terminal A+ current output +

5Terminal B pulse output -

6Terminal B+ Pulse output +

7PE terminal in housing

8PE terminal on connection piece between measuring sensor and signal converter.

Both grounding terminals 7 and 8 are equally effective from a technical point of view.

The following procedures are to be performed:

Unscrew the housing cover 1 of the electrical terminal compartment.

Thread the connection cable through the leadthrough in the housing.

Connect the cable for the current output and the cable for the optional pulse output as shown in the cable terminal diagrams below. To facilitate installation the connection plug can be removed from the device. The plug is configured in such a way as to prevent reverse polarity.

Connect the grounding to terminal 7 alternatively use the PE terminal on the connection piece between measuring sensor and signal converter 8.

Tighten the cable glands.

Hand-tight the housing cover with gasket.

32

www.siemens.com/flow

06/2009 - SFIDK.PS.050.F2.02 A5E02100423

Image 32
Contents Page Imprint Contents Contents Technical data Safety Instructions Intended useDevices are particularly suitable for the measurement Devices are rated for the following flow velocitiesCertifications CE markingDisclaimer Safety instructions from the manufacturerCopyright and data protection Product liability and warranty Information concerning the documentationSafety instructions for the operator Safety warnings are indicated by the following symbolsScope of delivery Instrument versionsDevices come in the following versions Following designs are available as optionsDevices with connection flange Sandwich versionFree air delivery measurement FAD optional Gross heat meter Dual seal Nameplate Storage TransportInstallation Installation condtitionsProhibited installation Measurement of liquidsRecommended installations for measurement of liquids Prohibited installation Measurement of vapours and gasesPreferred mounting position Pipelines with control valvePreferred mounting position Turning the connection housing Allen screws on connection housingTurning the display Turn the display as followsHeat insulation Inlet and outlet runs Minimum inlet runsMinimum outlet runs Flow straightenerInstallation General installation notesInstalling devices in flange design Installing devices in sandwich design Safety instructions Electrical ConnectionsFollowing procedures are to be performed Electrical ConnectionsElectrical connection of current and pulse output Maximum load resistance is calculated as followsPower supply Totalizer / pulse outputElectrical Connections Maximum frequency of the pulse output is fmax = 0.5 HzGrounding connections Protection category Therefore it is essential to observe the following pointsStart-up and control START-UPStart Operation Display and operating elementsSwitch from measuring mode to menu mode OperationOperating principles Functional description of the keysChanging the settings in the menu Navigation within the menu structureChanging units Measures in the event of faulty indications Totalizer / pulse outputMax. pulses per hour ExampleLevel Designation Explanation Volume Standard volume Mass Liquids, steams, gases Air GasOverview of the most important functions and units Volume Standard volume Mass Liquids, steams, gases Gas Temp Absolute Power Energy Density Pressure OverError messages Error message Cause MeasureMenu structure Overview of firmware versionsEntering values in change mode Character selection in change modeMenu item Quick Setup Test current is maintained until the menu has been exited Menu item TestsLevel Designation Selection / entry Explanation Menu item Setup firmware version basic Hart SV Operation Menu item Setup firmware version steam Reset heat energy Hart tertiary variable Temp. ↑ Pressure ↑ Density Operation Menu item Setup firmware version gas Hart SV RPM Pressure, temperature and density Operation Replacing signal converter / LCD display Following steps are to be performedService Availability of servicesSpare parts availability Returning the device to the manufacturerForm for copying to accompany a returned device Technical Data Functional principleMeasuring accuracy Technical dataMeasuring system Materials Installation conditionsOperating conditions Approvals and certifications Process connectionsElectrical connections Flange version EN Dimensions and weightsFlange versions Technical Data Flange version Asme B16.5 NPSSandwich version EN With Without Pressure sensor Size Pressure Dimensions inches Weight lbs RatingSandwich version Asme Flow tables Measuring rangesWater AirFt/s Flow rate limitsProduct Nominal sizes Minimum flow rates Maximum flow rates Measuring range saturated steam 1...7 bar Measuring range saturated steam 10.5...20 barKg/h 1092-1Lbs/h Measuring range saturated steam 15...100 psigMeasuring range saturated steam 150...300 psig Sitrans FX300