Siemens FX300 manual Changing units

Page 42

6 OPERATION

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SITRANS FX300

 

 

Example: changing the default parameter from m3/h to l/min

 

 

 

 

 

 

 

 

 

 

 

Procedure

Display

 

 

Procedure

Display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

107.2

 

5

 

^

0000600.00

 

 

 

 

 

 

m3/h

 

 

 

 

L/min

 

 

1

3 x

1.1.1

 

6

 

^

Display

 

 

 

 

 

 

Language

 

 

 

 

Unit

 

 

2

3 x

1.1.4

 

7

 

^

1.1.4

 

 

 

 

 

 

Max. Flow

 

 

 

 

Max. Flow

 

 

 

 

 

 

 

 

 

 

 

 

3

m3/h

 

8

 

^

Save Yes

 

 

 

 

 

 

Unit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4

3 x

L/min

 

9

 

^

1787.0000

 

 

 

 

 

 

Unit

 

 

 

 

L/min

 

 

 

 

 

 

 

 

 

 

 

 

 

6.2.5 Changing units

When entering numbers and values in floating point format, the maximum possible accuracy is 0.003%. The accuracy depends on both the position of the decimal point and the length of the number entered.

Numerical values and factors are displayed in the first line of the 10-digit display. Numerical values are either displayed in floating point format (123.4567890) or are expressed in exponential format (123456E002). The conversion factor of the totalizer and the pulse output is, however, expressed in whole numbers.

Input values exponential format

Display position

1

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

 

 

 

 

Decimal point

-

-

-

-

-

-

 

 

 

 

 

 

 

 

 

 

 

Input values

0 ...

0 ...

0 ...

0 ...

0 ...

0 ...

E

-

0 ...

0 ...

 

9

9

9

9

9

9

 

or

3

8

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

 

 

 

 

 

 

 

To select the exponential function, the decimal point must be located between the 2nd and 5th decimal position.

Flow units

The base units are m3/h for volume flow measurement, m3/h stand. for standard volumetric flow measurement and kg/h for mass flow measurement.

The flow units can be changed in menu item 1.1.4 "Max.Flow". User defined units can be entered using "User Def."

The unit (text) as well as the conversion factor (number) can be entered here. The conversion factor must always be entered based on the base unit.

Conversion

Formula

New unit (User Def.)

=

A1 Coeff.

*

Base unit

 

 

 

 

 

 

Example:

1 litre / h

=

0.001

*

m3/h

Menu items

1.1.3

Maes.Inst

Volume / Standard volume / Mass

 

 

 

1.1.4

Max. Flow

Unit (User Def.) / Text / A1 Coeff. / Max. Flow / Flow Displ

 

 

 

42

www.siemens.com/flow

06/2009 - SFIDK.PS.050.F2.02 A5E02100423

Image 42
Contents Page Imprint Contents Contents Technical data Devices are particularly suitable for the measurement Safety InstructionsIntended use Devices are rated for the following flow velocitiesCertifications CE markingSafety instructions from the manufacturer Copyright and data protectionDisclaimer Product liability and warranty Information concerning the documentationSafety instructions for the operator Safety warnings are indicated by the following symbolsDevices come in the following versions Scope of deliveryInstrument versions Following designs are available as optionsDevices with connection flange Sandwich versionFree air delivery measurement FAD optional Gross heat meter Dual seal Nameplate Storage TransportInstallation Installation condtitionsProhibited installation Measurement of liquidsRecommended installations for measurement of liquids Prohibited installation Measurement of vapours and gasesPipelines with control valve Preferred mounting positionPreferred mounting position Turning the connection housing Allen screws on connection housingTurning the display Turn the display as followsHeat insulation Inlet and outlet runs Minimum inlet runsMinimum outlet runs Flow straightenerInstallation General installation notesInstalling devices in flange design Installing devices in sandwich design Safety instructions Electrical ConnectionsFollowing procedures are to be performed Electrical ConnectionsElectrical connection of current and pulse output Maximum load resistance is calculated as followsPower supply Totalizer / pulse outputElectrical Connections Maximum frequency of the pulse output is fmax = 0.5 HzGrounding connections Protection category Therefore it is essential to observe the following pointsSTART-UP StartStart-up and control Operation Display and operating elementsOperating principles Switch from measuring mode to menu modeOperation Functional description of the keysChanging the settings in the menu Navigation within the menu structureChanging units Max. pulses per hour Measures in the event of faulty indicationsTotalizer / pulse output ExampleVolume Standard volume Mass Liquids, steams, gases Air Gas Overview of the most important functions and unitsLevel Designation Explanation Volume Standard volume Mass Liquids, steams, gases Gas Temp Absolute Power Energy Density Pressure OverError messages Error message Cause MeasureMenu structure Overview of firmware versionsEntering values in change mode Character selection in change modeMenu item Quick Setup Menu item Tests Level Designation Selection / entry ExplanationTest current is maintained until the menu has been exited Menu item Setup firmware version basic Hart SV Operation Menu item Setup firmware version steam Reset heat energy Hart tertiary variable Temp. ↑ Pressure ↑ Density Operation Menu item Setup firmware version gas Hart SV RPM Pressure, temperature and density Operation Replacing signal converter / LCD display Following steps are to be performedSpare parts availability ServiceAvailability of services Returning the device to the manufacturerForm for copying to accompany a returned device Technical Data Functional principleTechnical data Measuring systemMeasuring accuracy Installation conditions Operating conditionsMaterials Process connections Electrical connectionsApprovals and certifications Dimensions and weights Flange versionsFlange version EN Technical Data Flange version Asme B16.5 NPSSandwich version EN Size Pressure Dimensions inches Weight lbs Rating Sandwich version AsmeWith Without Pressure sensor Water Flow tablesMeasuring ranges AirFlow rate limits Product Nominal sizes Minimum flow rates Maximum flow ratesFt/s Kg/h Measuring range saturated steam 1...7 barMeasuring range saturated steam 10.5...20 bar 1092-1Measuring range saturated steam 15...100 psig Measuring range saturated steam 150...300 psigLbs/h Sitrans FX300