8 TECHNICAL DATA |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
| SITRANS FX300 |
| ||||
Flow rate limits |
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
| ||||||
Product | Nominal sizes | Minimum flow rates | Maximum flow rates |
|
| ||||||||
|
|
|
|
|
|
|
|
|
| ||||
| EN | ASME | [m/s] |
| [ft/s] | [m/s] |
| [ft/s] |
|
| |||
|
|
|
|
|
|
|
|
|
|
|
| ||
Liquids | DN15…300 | ½…12" | 0.5x(998/ | ρ 0.5 | 1 | 1.64x(998/ρ)0.5 2 | 7x(998/ | ρ 0.47 | 1 | 23x(998/ρ)0.47 2 |
|
| |
|
|
|
| ) |
| ) |
|
|
| ||||
Gas, steam | DN15…300 | ½…12" | 6x(1.29/ | ρ 0.5 | 3 | 16.4x(1.29/ρ)0.5 | 7x(998/ | ρ 0.47 | 3 | 23x(998/ρ)0.47 4 |
|
| |
|
|
|
| ) | 4 | ) |
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
| ρ = operating density [kg/m3] |
|
1Minimum flow rate 0.3 m/s - maximum flow rate 7 m/s
2Minimum flow rate 0.984 ft/s - maximum flow rate 23 ft/s
3Minimum flow rate 2 m/s - maximum flow rate 80 m/s; DN15 up to 45 m/s and DN25 up to 70 m/s
4Minimum flow rate 6.6 ft/s - maximum flow rate 262 ft/s; DN15 up to 148 ft/s and DN25 up to 230 ft/s
76 | www.siemens.com/flow | 06/2009 - SFIDK.PS.050.F2.02 A5E02100423 |