2.2.5 Remote Control Wiring (option) !
NOTE The following settings are available only for the model with remote options (R1).
(1)Turn the power switch off.
(2)Remove the transparent cover of the optional terminals.
(3)Connect the wires for remote control to the optional terminals. The “C” is the terminal common to other 1 to 5.
(4)Cover the terminals with the transparent cover and secure the screws.
Remote control terminals are arranged as follows:
|
| 2. |
|
| INSTALLATION |
REMOTE | ||
1 | 2 | 3 |
4 | 5 | C |
|
| ! |
Remote Control Terminal Arrangements |
WARNING
NOTE
Input characteristics: |
| |
Input signal | : Dry contact, | |
Control functions and signal types: |
| |
| • Recording start/stop | level |
| • Chart speed change | level |
| • Printout messages | trigger more than 250 msec |
Signal characteristics | : ON voltage | 0.5 V or less (30 mA DC) |
| Leakage current in OFF state | 0.25 mA or less |
| Signal duration | 250 msec or more |
Input types | : Photocoupler isolation (one line common) | |
| Internal isolated power source (5 V±5%) | |
Dielectric strength | : 500 V DC for one minute between input terminals and | |
| grounding terminal. |
|
To prevent an electric shock, ensure the main power switch is turned off before wiring.
•Use shielded cables for remote control signals to prevent noise pickup. The shield should be grounded at the recorder’s terminal.
•To prevent an emission of electromagnetic disturbances, separate the remote control wires from the power supply and input wires at least 0.1 m. Over 0.5 m is recommended.