6.2 Setting Offsets (OFFST)
6.2Setting Offsets (OFFST)
When recording and showing data using the current input as the reference, measured values can be corrected within 10% of the span. For example, assume the reading is 1.998 V for the current level of input. If you want to change the reading to 2.000 V, set 0.002 V as the offset. This menu opens only when offset settings protection is turned on in the SET UP mode.
Key operation
ESC
DISP
3 sec
CH UP
| ||
CH UP |
| |
FEED | ||
CH UP |
| |
| ESC |
|
CH UP | DISP | 3 sec |
|
|
Indication
Discription
Enter the SET mode by pressing the [DISP] key for three
seconds. Using the [ ] key, show [] and then press the [ENT] key.
Using the [ ] key, select the channel to be configured and then press the [ENT] key. Configurable channels are those whose measuring range is VOLT, TC, RTD or SCL.
Type the value of offset. Use the [ ] key to move through the digits and the [ ] key to select a numeral. When setting is complete, press the [ENT] key. Note that set the offset within 10% of the recording span or scale.
[] is displayed and the setting is then completed. Press the [ENT] key to set another channel; or press the [DISP] key for three seconds to return to the OPERATION mode.
NOTE If you press the [ESC] key in the middle of setting, the indication returns to the
[] and already set data will be aborted.