Omega RD260A instruction manual Difference Computation Setting Delt Setting

Page 58

6.1 Setting Input Range and Recording Span (RANGE)

Difference Computation Setting (DELT Setting)

To obtain the difference between the measured values of two channels, proceed as follows.

Key Operation

ESC

DISP

3 sec

PRINT

CH UP

PRINT

CH UP

PRINT

CH UP

PRINT

CH UP

Indication

Description

Press the [DISP] key for three seconds to enter the SET mode.

Select the [ ] display by using the [ ] key. Then press the [ENT] key.

Use the [ ] key to select the desired channel.

Note that channel No.1 cannot be of the ‘DELT’ type. Press the [ENT] key.

Use the [ ] key to select [

].

Press the [ENT] key.

 

The display which appears will allow you to enter a reference channel number. Note that the reference channel number must be lower than that of the channel being set. Therefore channel

No.1 cannot be of the [ ] type.

Use the [ ] key to select the desired reference channel number. Press the [ENT] key.

PRINT FEED

CH UP

PRINT FEED

CH UP

 

ESC

 

CH UP

DISP

3 sec

 

 

In case of VOLT the display which appears will show the same minimum value of the recording span as of the reference channel.

Use the [ ] and [ ] keys to select the minimum value (SPAN L).

Press the [ENT] key.

Note that in case of VOLT, the input range of the DELT-channel must be within the range of the reference channel. In case of TC, RTD, the minimum/maximum span values can be found by taking the ±value of the input range width. E.g. input range of TC type L = –200 to 900°C: minimum/maximum span values will be at ±1100 °C. The difference in measured value from the reference channel will be recorded on the DELT-channel.

The display which appears will show the same maximum value of the recording span as of the reference channel.

Use the [ ] and [ ] keys to select the maximum value (SPAN R) in the same way as described under ‘SPAN L setting’. Note that SPAN L cannot be equal to SPAN R. Press the [ENT] key.

[ ] is displayed and the setting is completed. Press the [ENT] key to set another channel; or press the [DISP] key for three seconds to return to the OPERATION mode.

6. SET MODE

NOTE If you press the [ESC] key in the middle of setting, the indication returns to the

[ ] and already set data will be aborted.

6 - 7

Image 58
Contents User’s Guide Servicing North America Introduction Power Supply Alternative currentDefects of Protective Grounding and Fuse Protective GroundingNecessity of Protective Grounding FuseInstallation ChaptersContents Setting A/D Integration Time Pen model Before Installation Handling PrecautionsChecking and Unpacking Checking Accessories and AppearanceAccessories No. Name Qty Remarks Spare Parts NameRemoving Packings Removing Lock Screw and Packings for TransitChecking Model Codes Checking Model CodesInstallation Site and Mounting Methods Installation SiteMounting Methods MountingExternal Dimensions Panel cutoutPen model WiringRear Panel Terminal Arrangement Dot-printing modelPower Supply Wiring Power Supply WiringInput Signal Wiring Crimp-On LugsRTD input Wiring for Input TerminalsTC input Terminal BoardOutput type Alarm Output Wiring optionContact characteristics Output capacityRemote Control Wiring option Input characteristicsComponent Names and Functions Front Panel Power switch push-button typeKey panel push-key type Status Display Chart cassetteData Display Characters on the Display Maintenance Switching Power On/OffInstalling or Replacing Chart Paper Prepare a new chart Replacement under pen-replacement mode Installing or Replacing Pens Pen modelSimple replacement Felt-tip penPlotter pen Installing or Replacing Ribbon Cassette Dot-printing model Starting/Stoping the Recording Feeding the ChartGetting the Recording Printing SamplesGetting a Digital Printout Getting a List PrintoutGetting a SET UP List Printout Dot-printing model Printing SampleDisplay MAN except for the 1-pen model Selectting the DisplaySelecting a screen Periodic Maintenance MaintenanceIndication of Remaining Chart Length Cleaning the Carriage Shafts Battery ReplacementFuse Replacement Fuse ReplacementPen model CalibrationDot-printing model Required instruments for calibration Calibration procedureDC Voltage Signal for Pen model same as Dot-printing model Recommended instrumentsTemperature Compensation for Thermocouple Signal Reference Junction Compensation for Thermocouple InputPen-Position Adjustment Pen model Adjusting procedureDaily OPERATION/MAINTENANCE Dot-Position Adjustment Dot-printing model Figure aDaily OPERATION/MAINTENANCE Data Setting Guides SET UP mode Operation modeSET mode Turning the function offOperation Mode Disp keySET Mode SET UP Mode Password Procedures to Enter Setting Data SET Mode Setting Input Range and Recording Span Range Key operation Skip SettingIndication Description Voltage Measurement Setting Volt Setting Volt Input Ranges DisplayKey operation Indication Description TC Type Description and RTD Type DescriptionDigital Input Setting DI Setting Difference Computation Setting Delt Setting Scale Setting SCL Setting Indication DiscriptionKey operation Indication Discription Setting to Obtain Square Root Sqrt SettingENT key Explanation of Square Root Setting Offsets Offst Setting Alarms Alarm Is displayed and the setting is completed. Press Assigning Units Unit Setting Chart Speed Chart Chart Speed unit mm/hSetting Date/Time Clock FIX Auto Assigning Tags TAG Setting Messages MSG Means ‘SPACE’Setting Key-lock KEY SET UP Mode Setting Procedure Setting Alarm Hysteresis On/Off Setting A/D Integration Time Pen model Setting Upscale/Downscale TC Burnout Select Using the key. Press the ENT key Display AppearsConfiguring the Input Filter Pen Model Filter works as followsBuffer data for Setting for Moving Average Calculations Dot-printing modelNth sample +1th sample Turning Offset Settings Protection On/Off Assigning Different Colors to Channels Dot-printing model To select Display, and press the ENT key SelectSetting Pen Offset Compensation On/Off Pen model, /D1 option Recorder as Viewed from the SideTurning Alarm Printing On/Off Use Key to select the display Press the ENT keyPen Model Chart Speed Setting Periodic Printout On/OffDot-printing Model Chart Speed Turning Printing of Changes in Chart Speed On/Off Setting ProcedureSetting Procedure Selecting channel Number or Tag Changing the Temperature Unit /D2 option Select Using the key Press the ENT key. The display AppearsChanging Default Settings for Remote Operation option Turning the Settings Protection Function On/Off Initializing Settings Error Codes Error Code ListRecovery Self-diagnostic Message Printout Meanings Store SET UP Settings AgaiinTroubleshooting Recorder is completely inoperativeIs the input free of noise? Glitch is found other than the previously mentioned ploblems Warrant Y/ Disclaimer OMEGA…Of Course