8.2 Troubleshooting
8.2 Troubleshooting
Recorder is completely inoperative
Is the power switch ON?
Does the switch turn properly?
YES
Is the power supply properly connected?
YES
Does the power meet the specified rating?
YES
Is the fuse blown?
YES
Replace the fuse.
YES
Is the fuse blown just after replaced?
YES
Contact OMEGA Engineering.
NO | See section 4.1. |
Turn the switch on properly |
NO | See subsection 2.2.2. |
Connect the supply lines properly. |
NO | Supply the power with specified |
| voltage/frequency. |
NO | Contact OMEGA Engineering. |
See subsection 4.8.3.
NO
End
8. TROUBLESHOOTING
8 - 3