7.3 Setting A/D Integration Time (Pen model)

The purpose of the following setting is to specify which A/D integration time will be used. Depending on the power supply, the integration time will affect the ability of noise rejection and therefore the recording.

If the power supply of the recorder and of the measuring object are the same, the effect of noise will be automatically minimized and the (initial) value of AUTO is the recommended one.

If the power supplies are different, you should select from 50Hz or 60Hz integration time. The choice depends on the power supply of the instrument with the highest noise level.

There are three possibilities:

AUTO: integration time depends on the power supply of the recorder.

50Hz

60Hz.

The initial value is ‘AUTO’.

SETTING PROCEDURE:

Enter the SET UP mode by turning ‘ON’ the power while pressing and holding the [CH UP] (ENT) key until the SET UP display appears.

Use the [ ] key to select the display []. Press the [ENT] key.

Select ‘AUTO’, ‘50’ or ‘60’ according to your situation using the [ ] key.

Then press the [ENT] key. The display [] appears.

Before leaving the SET UP mode, you have to store your new settings. Use the [] key

to select the [] display, and press the [ENT] key.

Select [] to keep your new settings or [] to abort the settings, and press the [ENT] key. After a few seconds, the OPERATION mode will appear.

7 - 4

Image 75
Contents User’s Guide Servicing North America Introduction Alternative current Power SupplyFuse Protective GroundingNecessity of Protective Grounding Defects of Protective Grounding and FuseChapters InstallationContents Setting A/D Integration Time Pen model Handling Precautions Before InstallationSpare Parts Name Checking Accessories and AppearanceAccessories No. Name Qty Remarks Checking and UnpackingRemoving Lock Screw and Packings for Transit Removing PackingsChecking Model Codes Checking Model CodesInstallation Site Installation Site and Mounting MethodsMounting Mounting MethodsPanel cutout External DimensionsDot-printing model WiringRear Panel Terminal Arrangement Pen modelPower Supply Wiring Power Supply WiringCrimp-On Lugs Input Signal WiringTerminal Board Wiring for Input TerminalsTC input RTD inputOutput capacity Alarm Output Wiring optionContact characteristics Output typeInput characteristics Remote Control Wiring optionComponent Names and Functions Power switch push-button type Front PanelKey panel push-key type Chart cassette Status DisplayData Display Characters on the Display Switching Power On/Off MaintenanceInstalling or Replacing Chart Paper Prepare a new chart Felt-tip pen Installing or Replacing Pens Pen modelSimple replacement Replacement under pen-replacement modePlotter pen Installing or Replacing Ribbon Cassette Dot-printing model Printing Samples Feeding the ChartGetting the Recording Starting/Stoping the RecordingGetting a List Printout Getting a Digital PrintoutDot-printing model Printing Sample Getting a SET UP List PrintoutSelectting the Display Display MAN except for the 1-pen modelSelecting a screen Maintenance Periodic MaintenanceIndication of Remaining Chart Length Fuse Replacement Battery ReplacementFuse Replacement Cleaning the Carriage ShaftsCalibration Pen modelDot-printing model Recommended instruments Calibration procedureDC Voltage Signal for Pen model same as Dot-printing model Required instruments for calibrationReference Junction Compensation for Thermocouple Input Temperature Compensation for Thermocouple SignalAdjusting procedure Pen-Position Adjustment Pen modelDaily OPERATION/MAINTENANCE Figure a Dot-Position Adjustment Dot-printing modelDaily OPERATION/MAINTENANCE Data Setting Guides Turning the function off Operation modeSET mode SET UP modeDisp key Operation ModeSET Mode SET UP Mode Password Procedures to Enter Setting Data SET Mode Setting Input Range and Recording Span Range Skip Setting Key operationIndication Description Volt Input Ranges Display Voltage Measurement Setting Volt SettingTC Type Description and RTD Type Description Key operation Indication DescriptionDigital Input Setting DI Setting Difference Computation Setting Delt Setting Indication Discription Scale Setting SCL SettingSetting to Obtain Square Root Sqrt Setting Key operation Indication DiscriptionENT key Explanation of Square Root Setting Offsets Offst Setting Alarms Alarm Is displayed and the setting is completed. Press Assigning Units Unit Chart Speed unit mm/h Setting Chart Speed ChartSetting Date/Time Clock FIX Auto Assigning Tags TAG Means ‘SPACE’ Setting Messages MSGSetting Key-lock KEY SET UP Mode Setting Procedure Setting Alarm Hysteresis On/Off Setting A/D Integration Time Pen model Select Using the key. Press the ENT key Display Appears Setting Upscale/Downscale TC BurnoutFilter works as follows Configuring the Input Filter Pen ModelSetting for Moving Average Calculations Dot-printing model Buffer data forNth sample +1th sample Turning Offset Settings Protection On/Off To select Display, and press the ENT key Select Assigning Different Colors to Channels Dot-printing modelRecorder as Viewed from the Side Setting Pen Offset Compensation On/Off Pen model, /D1 optionUse Key to select the display Press the ENT key Turning Alarm Printing On/OffSetting Periodic Printout On/Off Pen Model Chart SpeedDot-printing Model Chart Speed Setting Procedure Turning Printing of Changes in Chart Speed On/OffSetting Procedure Selecting channel Number or Tag Select Using the key Press the ENT key. The display Appears Changing the Temperature Unit /D2 optionChanging Default Settings for Remote Operation option Turning the Settings Protection Function On/Off Initializing Settings Error Code List Error CodesRecovery Store SET UP Settings Agaiin Self-diagnostic Message Printout MeaningsRecorder is completely inoperative TroubleshootingIs the input free of noise? Glitch is found other than the previously mentioned ploblems Warrant Y/ Disclaimer OMEGA…Of Course