Omega RD260A instruction manual Scale Setting SCL Setting, Indication Discription

Page 59

Scale Setting (SCL Setting)

Key operation

ESC

DISP

3 sec

PRINT

CH UP

PRINT

CH UP

PRINT

CH UP

PRINT

CH UP

PRINT FEED

CH UP

PRINT FEED

CH UP

PRINT FEED

CH UP

PRINT FEED

CH UP

 

ESC

 

CH UP

DISP

3 sec

 

 

NOTE

To assign a different scale to the measured data, the range can be set as described below. Note that the measured data which can get a different scale must be of the voltage (VOLT). To assign a unit to this new scale, see section 6.4.

Indication

Discription

Press the [DISP] key for three seconds to enter the SET mode.

Select the [ ] display by using the [ ] key.

Then press the [ENT] key.

Use the [ ] key to select the desired channel.

Press the [ENT] key.

Use the [ ] key to select [ ].

Press the [ENT] key.

Select the desired range (using the [ ] key) and press the [ENT] key.

The display which appears will allow you to enter the minimum value of the recording span (SPAN L). Select the desired minimum value using the [ ] and [ ] keys. Then press the [ENT] key.

The display which appears will allow you to enter the maximum value of the recording span (SPAN R). Select it the same way as the minimum value and press the [ENT] key. Note that SPAN L and SPAN R cannot be same.

This display allows you to enter the minimum value of the scale (SCL L) corresponding to the minimum value of the recording span (SPAN L). Select it using the [ ] and the [ ] keys. The position of the decimal point can be moved using the [ ] key in case [ ] is flashing. Note that the allowable range is –19999 to 20000. Then press the [ENT] key.

This display allows you to enter the desired maximum value of the scale (SCL R) corresponding to the maximum value of the recording span (SPAN R). Select it the same way as the minimum value. Then press the [ENT] key. Note that the decimal point will be in the same position as for SCL L. SCL L and SCL R cannot be same.

[ ] is displayed and the setting is completed. Press the [ENT] key to set another channel; or press the [DISP] key for three seconds to return to the OPERATION mode.

In this example, 0.000V input will be scaled to –100.0. 2.000V input will be scaled to 100.0.

If you press the [ESC] key in the middle of setting, the indication returns to the

[ ] and already set data will be aborted.

6 - 8

Image 59
Contents User’s Guide Servicing North America Introduction Alternative current Power SupplyFuse Protective GroundingNecessity of Protective Grounding Defects of Protective Grounding and FuseChapters InstallationContents Setting A/D Integration Time Pen model Handling Precautions Before InstallationSpare Parts Name Checking Accessories and AppearanceAccessories No. Name Qty Remarks Checking and UnpackingRemoving Lock Screw and Packings for Transit Removing PackingsChecking Model Codes Checking Model CodesInstallation Site Installation Site and Mounting MethodsMounting Mounting MethodsPanel cutout External DimensionsDot-printing model WiringRear Panel Terminal Arrangement Pen modelPower Supply Wiring Power Supply WiringCrimp-On Lugs Input Signal WiringTerminal Board Wiring for Input TerminalsTC input RTD inputOutput capacity Alarm Output Wiring optionContact characteristics Output typeInput characteristics Remote Control Wiring optionComponent Names and Functions Key panel push-key type Power switch push-button typeFront Panel Data Display Chart cassetteStatus Display Characters on the Display Switching Power On/Off MaintenanceInstalling or Replacing Chart Paper Prepare a new chart Felt-tip pen Installing or Replacing Pens Pen modelSimple replacement Replacement under pen-replacement modePlotter pen Installing or Replacing Ribbon Cassette Dot-printing model Printing Samples Feeding the ChartGetting the Recording Starting/Stoping the RecordingGetting a List Printout Getting a Digital PrintoutDot-printing model Printing Sample Getting a SET UP List PrintoutSelecting a screen Selectting the DisplayDisplay MAN except for the 1-pen model Indication of Remaining Chart Length MaintenancePeriodic Maintenance Fuse Replacement Battery ReplacementFuse Replacement Cleaning the Carriage ShaftsDot-printing model CalibrationPen model Recommended instruments Calibration procedureDC Voltage Signal for Pen model same as Dot-printing model Required instruments for calibrationReference Junction Compensation for Thermocouple Input Temperature Compensation for Thermocouple SignalAdjusting procedure Pen-Position Adjustment Pen modelDaily OPERATION/MAINTENANCE Figure a Dot-Position Adjustment Dot-printing modelDaily OPERATION/MAINTENANCE Data Setting Guides Turning the function off Operation modeSET mode SET UP modeDisp key Operation ModeSET Mode SET UP Mode Password Procedures to Enter Setting Data SET Mode Setting Input Range and Recording Span Range Indication Description Skip SettingKey operation Volt Input Ranges Display Voltage Measurement Setting Volt SettingTC Type Description and RTD Type Description Key operation Indication DescriptionDigital Input Setting DI Setting Difference Computation Setting Delt Setting Indication Discription Scale Setting SCL SettingENT key Setting to Obtain Square Root Sqrt SettingKey operation Indication Discription Explanation of Square Root Setting Offsets Offst Setting Alarms Alarm Is displayed and the setting is completed. Press Assigning Units Unit Chart Speed unit mm/h Setting Chart Speed ChartSetting Date/Time Clock FIX Auto Assigning Tags TAG Means ‘SPACE’ Setting Messages MSGSetting Key-lock KEY SET UP Mode Setting Procedure Setting Alarm Hysteresis On/Off Setting A/D Integration Time Pen model Select Using the key. Press the ENT key Display Appears Setting Upscale/Downscale TC BurnoutFilter works as follows Configuring the Input Filter Pen ModelNth sample +1th sample Setting for Moving Average Calculations Dot-printing modelBuffer data for Turning Offset Settings Protection On/Off To select Display, and press the ENT key Select Assigning Different Colors to Channels Dot-printing modelRecorder as Viewed from the Side Setting Pen Offset Compensation On/Off Pen model, /D1 optionUse Key to select the display Press the ENT key Turning Alarm Printing On/OffDot-printing Model Chart Speed Setting Periodic Printout On/OffPen Model Chart Speed Setting Procedure Turning Printing of Changes in Chart Speed On/OffSetting Procedure Selecting channel Number or Tag Select Using the key Press the ENT key. The display Appears Changing the Temperature Unit /D2 optionChanging Default Settings for Remote Operation option Turning the Settings Protection Function On/Off Initializing Settings Recovery Error Code ListError Codes Store SET UP Settings Agaiin Self-diagnostic Message Printout MeaningsRecorder is completely inoperative TroubleshootingIs the input free of noise? Glitch is found other than the previously mentioned ploblems Warrant Y/ Disclaimer OMEGA…Of Course