Omega RD260A instruction manual Installing or Replacing Chart Paper

Page 27

4.2Installing (or Replacing) Chart Paper

(1)Open the front door of the recorder.

(2)Check the recorder is not in recording state; but the power ON is allowed.

(3)Riffle the chart thoroughly before loading.

(4)While gently pressing the chart-cassette stoppers, located at both front ends of the cassette, lift and draw the cassette out of the recorder case.

Stoppers

(5)Open the chart retainer backward on the cassette.

(6)Pull and open the front transparent chart guide.

Chart retainer

Chart guide (transparent plastic plate)

(7)Load the chart paper into the chart compartment; the small perforations should be at the left-hand. Make sure that the sprocket teeth of the chart drives are properly engaged in the chart paper perforations. Take care not to load the chart paper backwards.

Small perforations

4 - 2

Image 27
Contents User’s Guide Servicing North America Introduction Alternative current Power SupplyFuse Protective GroundingNecessity of Protective Grounding Defects of Protective Grounding and FuseChapters InstallationContents Setting A/D Integration Time Pen model Handling Precautions Before InstallationSpare Parts Name Checking Accessories and AppearanceAccessories No. Name Qty Remarks Checking and UnpackingRemoving Lock Screw and Packings for Transit Removing PackingsChecking Model Codes Checking Model CodesInstallation Site Installation Site and Mounting MethodsMounting Mounting MethodsPanel cutout External DimensionsDot-printing model WiringRear Panel Terminal Arrangement Pen modelPower Supply Wiring Power Supply WiringCrimp-On Lugs Input Signal WiringTerminal Board Wiring for Input TerminalsTC input RTD inputOutput capacity Alarm Output Wiring optionContact characteristics Output typeInput characteristics Remote Control Wiring optionComponent Names and Functions Power switch push-button type Front PanelKey panel push-key type Chart cassette Status Display Data Display Characters on the Display Switching Power On/Off MaintenanceInstalling or Replacing Chart Paper Prepare a new chart Felt-tip pen Installing or Replacing Pens Pen modelSimple replacement Replacement under pen-replacement modePlotter pen Installing or Replacing Ribbon Cassette Dot-printing model Printing Samples Feeding the ChartGetting the Recording Starting/Stoping the RecordingGetting a List Printout Getting a Digital PrintoutDot-printing model Printing Sample Getting a SET UP List PrintoutSelectting the Display Display MAN except for the 1-pen modelSelecting a screen Maintenance Periodic MaintenanceIndication of Remaining Chart Length Fuse Replacement Battery ReplacementFuse Replacement Cleaning the Carriage ShaftsCalibration Pen modelDot-printing model Recommended instruments Calibration procedureDC Voltage Signal for Pen model same as Dot-printing model Required instruments for calibrationReference Junction Compensation for Thermocouple Input Temperature Compensation for Thermocouple SignalAdjusting procedure Pen-Position Adjustment Pen modelDaily OPERATION/MAINTENANCE Figure a Dot-Position Adjustment Dot-printing modelDaily OPERATION/MAINTENANCE Data Setting Guides Turning the function off Operation modeSET mode SET UP modeDisp key Operation ModeSET Mode SET UP Mode Password Procedures to Enter Setting Data SET Mode Setting Input Range and Recording Span Range Skip Setting Key operationIndication Description Volt Input Ranges Display Voltage Measurement Setting Volt SettingTC Type Description and RTD Type Description Key operation Indication DescriptionDigital Input Setting DI Setting Difference Computation Setting Delt Setting Indication Discription Scale Setting SCL SettingSetting to Obtain Square Root Sqrt Setting Key operation Indication DiscriptionENT key Explanation of Square Root Setting Offsets Offst Setting Alarms Alarm Is displayed and the setting is completed. Press Assigning Units Unit Chart Speed unit mm/h Setting Chart Speed ChartSetting Date/Time Clock FIX Auto Assigning Tags TAG Means ‘SPACE’ Setting Messages MSGSetting Key-lock KEY SET UP Mode Setting Procedure Setting Alarm Hysteresis On/Off Setting A/D Integration Time Pen model Select Using the key. Press the ENT key Display Appears Setting Upscale/Downscale TC BurnoutFilter works as follows Configuring the Input Filter Pen ModelSetting for Moving Average Calculations Dot-printing model Buffer data forNth sample +1th sample Turning Offset Settings Protection On/Off To select Display, and press the ENT key Select Assigning Different Colors to Channels Dot-printing modelRecorder as Viewed from the Side Setting Pen Offset Compensation On/Off Pen model, /D1 optionUse Key to select the display Press the ENT key Turning Alarm Printing On/OffSetting Periodic Printout On/Off Pen Model Chart SpeedDot-printing Model Chart Speed Setting Procedure Turning Printing of Changes in Chart Speed On/OffSetting Procedure Selecting channel Number or Tag Select Using the key Press the ENT key. The display Appears Changing the Temperature Unit /D2 optionChanging Default Settings for Remote Operation option Turning the Settings Protection Function On/Off Initializing Settings Error Code List Error CodesRecovery Store SET UP Settings Agaiin Self-diagnostic Message Printout MeaningsRecorder is completely inoperative TroubleshootingIs the input free of noise? Glitch is found other than the previously mentioned ploblems Warrant Y/ Disclaimer OMEGA…Of Course