2.2.3Input Signal Wiring !
(1)Turn off the recorder power switch.
(2)Remove the transparent cover for the input signal terminals.
(3)Connect the input signal wires to the terminals.
(4)Attach the cover on the terminals, and secure the screws.
It is recommended to use the
WARNING
CAUTION
NOTE
Crimp-On Lugs
Suggestions to minimize noise pickup:
•Separate the
•It is desirable that the measuring object is not a noise source; if not, insulate it from the
•It is recommended to use shielded wires to minimize noise pickup from an electrostatic induction source. The shields are connected to the recorder’s grounding terminal, if necessary. In that case, avoid
•To minimize the noize from an electrostatic induction noise, twist the measuring wires (a pair of wires) in short and equal intervals.
•The grounding line must have low resistance.
In case of using a thermocouple (TC), the recorder is equipped with a reference junction compensator, so the ambient temperature around the recorder’s terminal board should be kept as stable as possible.
•Cover the input terminals with the transparent cover.
•Do not use a
•Keep the ambient temperature as stable as possible; start/stop of a nearby fan affects the temperature.
If input signal wires are connected to two or more receivers with parallel connection, measured values sometimes affect each other. The following notes should be observed for parallel connection:
•Ground each receiver at the same point.
•Do not turn on/off the power of a receiver while other receivers are in measuring operation.
•A resistance temperature detector (RTD) cannot be used essentially in parallel connection.
To prevent an electric shock, ensure the main power switch is turned off when wiring.
Do not apply inputs exceeding the rated range as follows to prevent the recorder from damage:
•Maximum input voltage
±10 V DC for TC, and for the voltage range of 2 V DC or less ±30 V DC for the voltage range of 6 to 20 V DC
•Maximum common mode voltage 250 V AC rms (50/60 Hz)
•The installation of each input is Category II.
•To prevent an emission of electromagnetic disturbances, separate the input wires from the other wires at least 0.1 m. Over 0.5 m is recommended.
2 - 6