Teledyne AX300-I manual Sensor Installation or Replacement, Installing the R17MED Sensor

Page 15

Portable Oxygen Analyzer

Operation

 

 

 

2.1.1 Sensor Installation or Replacement

Note: The R17MED oxygen sensor must be installed before the oxygen analyzer/analyzer can be operated

1.Remove the new sensor from its protective bag. Inspect the sensor for damage or electrolyte leakage. If the sensor is damaged, obtain a replacement. Do not use the defective sensor as it may damage the unit.

WARNING: THE SENSOR ELECTROLYTE IS CAUSTIC. DO NOT LET IT COME IN CONTACT WITH SKIN. IF IT DOES, FLUSH AFFECTED AREA WITH WATER. DO NOT ATTEMPT TO OPEN OR REPAIR THE SENSOR.

WARNING: THE SENSOR ALSO CONTAINS LEAD. LEAKING OR EXHAUSTED SENSORS SHOULD BE HANDLED AND DISPOSED OF IN ACCORDANCE WITH LOCAL REGULATIONS. CONTACT TELEDYNE FOR THE MATERIAL SAFETY DATA SHEET

2.Plug one end of the coiled cable into the jack receptacle on the back end of the R17MED sensor and secure in place with the capture nut located at the base of the connector. See Figure 2-1.

Figure 2-1: Installing the R17MED Sensor

3.Plug the other end of the coiled cable into the receptacle located on the right side of the unit and secure it in place using the capture nut. See Figure 2-2.

Teledyne Analytical Instruments

15

Image 15
Contents AX300-I AX300-I About This Manual How To Use This ManualSafety Messages Table of Contents Index Teledyne Analytical Instruments AX300 Front View List of FiguresList of Tables AX300 Front View IntroductionApplicable Standards FeaturesOptions Digit LCD display. For RS-232 reconfiguration, see Section Applications Theory of Operation Sensor Signal Processing Operation SetupInstalling the R17MED Sensor Sensor Installation or ReplacementSensor Cable Connection to Analyzer MountingV-Mount Adapter Installation Battery InstallationInstalling Batteries Calibration S CalibrationOperation AX300-I 0-1 VDC or RS 232 Digital Output Port Output 0-1 VDC or RS232Use ProcedurePortable Oxygen Analyzer Operation Pressure TemperatureGas Sampling HumidityAnesthetic Gases Discrepancy in ReadingsOperation AX300-I Oxygen Reading Error in a Mixture of Anesthetic Gas Do’s and Don’ts CleaningPortable Oxygen Analyzer Operation Blank General Service Information Overall MaintenanceBattery Maintenance Calibration Sensor MaintenanceHumidity Discrepancy in Readings Oxygen Reading Error Troubleshooting Troubleshooting√ Sensor Concentration of O2 is Watchdog Error CodesOther Problems with the Instrument Return Authorization for Service SpecificationsAppendix Appendix AX300-I Optional Accessories Spare Parts ListInstrument for which the parts are intended Blank Index Index