Portable Oxygen Analyzer | Service Manual | |||
|
|
|
| |
|
|
| ||
|
| adapter to prevent back filling. | ||
|
| O2 flow rate should not | ||
|
| exceed 5 l/min. | ||
|
| C) Oxygen concentration at the | ||
|
| sensor is significantly higher | ||
|
| than 21%. Take the | ||
|
| instrument to a well- | ||
|
| ventilated area and check | ||
|
| the reading again. | ||
|
| E) Try calibrating with a known | ||
|
| good sensor; if this fails, see | ||
|
| symptom "Reading drifts | ||
|
| over | ||
The sensor does not | Water is condensing | A) Remove the sensor from tee | ||
react to changes in | on the sensing | adapter and unscrew the | ||
oxygen concentration, | surface. | plastic flow diverter. Using | ||
or the readings are | Electrical interference | absorbent tissue or cotton | ||
unstable and drifting. | swab, gently wipe off sensing | |||
is disrupting the | ||||
| surface inside threaded | |||
| electronics | |||
| portion of sensor assembly. | |||
|
| |||
|
| B) Relocate unit away from | ||
|
| sources of electrical noise | ||
|
| such as cauterizing equipment | ||
|
| and | ||
|
|
| ||
The display is flashing | The unit has detected | A) Check sensor cable | ||
√ SENSOR | a fault in the signal | connections and make sure | ||
from the sensor. | they are completely | |||
| ||||
| Sensor has expired. | inserted into the mating | ||
| connector and the capture | |||
|
| |||
| The sensor has been | nut is firmly in place. | ||
| exposed to a gas | B) Expose the sensor in 100% | ||
| containing little or no | |||
| O2 and check calibration. | |||
| oxygen. | |||
| C) The sensor output has | |||
|
| |||
|
| fallen to a level where it is | ||
|
| no longer usable. Replace | ||
|
| sensor. | ||
The oxygen reading | Like all O2 sensors, | A) During calibration, make | ||
fluctuates or appears to | the R17MED detects | sure there is no restrictions | ||
be incorrect. | the changes in the | on exhaust side of sensor. If | ||
| partial pressure of | the reading changes with | ||
| O2. | flow, the sensor is | ||
|
| pressurized or there may be | ||
|
| a leak in the system. | ||
|
| B) If a high degree of accuracy | ||
|
| is desired, or the |
Teledyne Analytical Instruments | 37 |