Service Manual |
|
| |
|
|
|
|
|
|
| |
|
| concentration of O2 is in | |
|
| excess of 40%, calibration | |
|
| with 100% is recommended. | |
|
| C) If humidified gas is used to | |
|
| ventilate the patient, water | |
|
| vapor actually dilutes the | |
|
| gas. See Appendix: | |
|
| Humidity, Temperature. | |
|
| D) If a blender is used, check | |
|
| its calibration. See | |
|
| Appendix: Discrepancy in | |
|
| Readings. | |
The unit has stopped | The | A) Disconnect the batteries and | |
working and the LCD is | equipped with an | inspect the contacts for | |
displaying | electronic "watch dog," | corrosion. Reconnect the | |
alphanumeric figures. | which analyzers the | batteries. If the unit functions | |
| circuitry within the unit | properly, calibrate the unit | |
| for potential faults and | and reset the alarm values. | |
| renders the unit | B) Try a new set of batteries. | |
| inoperable until the | ||
|
|
| |
| condition is corrected. | C) Increase the distance | |
| (See Watch dog | between the unit and any | |
| section below) | source of radio frequency | |
| Several conditions can | interference. The sensor | |
| cable is a prime source of | ||
| activate the "watch | ||
| pickup as it can act like an | ||
| dog." Dropping the | ||
| antenna. Relocate the | ||
| unit, poor battery | ||
| sensor cable and if possible | ||
| connections, and radio | ||
| change its coiled length to | ||
| frequency interference | ||
| |||
| are the most common | ||
| Placing the cable in a | ||
| causes. See the watch | ||
| different position may also | ||
| dog section for | ||
| help. | ||
| additional information | ||
|
|
| |
No display. | A) Batteries expired. | A) Check/replace batteries. | |
| B) Bad battery | B) Check battery connections. | |
| connection. | C) Calibrate. | |
|
| ||
|
|
| |
Keys inoperable/cannot | The LOCK/UNLOCK | A) Press LOCK/UNLOCK key | |
turn unit off | key is activated which | once. LCD will flash | |
| is preventing key | indicating keys are active. | |
| operation |
|
|
Cannot adjust | Critical settings require | A) If display is not flashing | |
calibration | two keys be pressed in | press lock key once to | |
| a specific order. | activate keys. Press desired | |
| LOCK/UNLOCK key is | function followed by the Up | |
| and Down key. | ||
| active | ||
|
|
|
Teledyne Analytical Instruments | 38 |