Teledyne AX300-I manual Cleaning, Do’s and Don’ts

Page 28

Operation

AX300-I

 

 

 

oxygen reading continues to drop after each use in nitrous oxide the sensor should be removed from service. If the sensor can no longer be calibrated or if there is any sign of electrolyte leakage, the sensor should be disposed of in accordance with local regulations and the Material Safety Data Sheet (MSDS) available through Teledyne.

2.3.6 Cleaning

The AX300-I and R17MED sensor are non-sterile devices.

The oxygen analyzer, oxygen sensor, and sensor interconnection cable may be cleaned by wiping the surfaces with isopropyl alcohol or a mild cleaning solution.

Before placing instrument in operation make sure all surface are dry and unit is properly calibrated.

When cleaning do not allow liquids or moisture to enter the instrument or sensor internal cavities. Do not allow the cleaning solution to come in contact with the electrical connections. Do not immerse the instrument, oxygen sensor or sensor interconnection cable in water or any other liquid.

Do not expose the instrument, oxygen sensor and interconnection cable to steam, ethylene oxide, or radiation sterilization.

2.4 Do’s and Don’ts

– DO –

Read all of the directions before using for the first time.

Calibrate every 8 hours or before every use.

Visually inspect the sensor for leakage before each use.

Calibrate using 100% oxygen and check in air.

Test batteries regularly and replace when battery indicator shows low battery (no bars remaining)

Make sure keys are locked by using the LOCK/UNLOCK key feature.

Keep the unit, sensor and connections dry, or on the dry side of the breathing circuit.

Recalibrate after replacing the batteries.

Teledyne Analytical Instruments

28

Image 28
Contents AX300-I AX300-I How To Use This Manual About This ManualSafety Messages Table of Contents Index Teledyne Analytical Instruments List of Figures AX300 Front ViewList of Tables Introduction AX300 Front ViewFeatures Applicable StandardsDigit LCD display. For RS-232 reconfiguration, see Section OptionsTheory of Operation ApplicationsSensor Signal Processing Setup OperationSensor Installation or Replacement Installing the R17MED SensorMounting Sensor Cable Connection to AnalyzerBattery Installation V-Mount Adapter InstallationInstalling Batteries Calibration Calibration SOperation AX300-I Output 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output PortProcedure UsePortable Oxygen Analyzer Operation Temperature Gas SamplingHumidity Pressure Discrepancy in Readings Anesthetic GasesOperation AX300-I Oxygen Reading Error in a Mixture of Anesthetic Gas Cleaning Do’s and Don’tsPortable Oxygen Analyzer Operation Blank Overall Maintenance General Service InformationBattery Maintenance Sensor Maintenance CalibrationHumidity Discrepancy in Readings Oxygen Reading Error Troubleshooting Troubleshooting√ Sensor Concentration of O2 is Error Codes WatchdogOther Problems with the Instrument Specifications Return Authorization for ServiceAppendix Appendix AX300-I Spare Parts List Optional AccessoriesInstrument for which the parts are intended Blank Index Index