Teledyne AX300-I manual Discrepancy in Readings

Page 34

Service Manual

AX300-I

 

 

 

When the sensor is connected to a ventilator circuit, the alternating “breathing” pressure cycles generated by the ventilator will be sensed as an increase in the oxygen percentage (especially if the sensor is fast enough to sense the changes, as is the R17MED). In reality, the percentage of oxygen is not changing; it is the total pressure that is increasing producing a corresponding increase in the partial pressure of oxygen. A hundred centimeter of water pressure pulse will produce a 0.11 atmosphere, or an 11% increase in the total and therefore partial pressure of oxygen. Assuming that the sensor is fast enough to track this pressure pulse, an un-pressurized reading of 50% oxygen will increase to 55.3% if the sensor is subjected to a pressure cycle of 100cm H2O. The reading will rise proportionally less for smaller pressures.

3.6.4 Discrepancy in Readings

The AX300-I instrument should be used to measure the oxygen concentration exiting another oxygen mixing device or life support system (i.e., a blender, incubator or anesthesia machine). The information obtained from the AX300-I should never be used to adjust a life-support system, but should only be used as an indication that the life support system or device may require service and/or calibration.

When a discrepancy in oxygen readings is detected, the oxygen analyzers readings should be verified by checking the AX300-I battery condition and calibration using 100% oxygen. If the analyzer can be calibrated, the unit can be assumed to be in good working order and capable of providing readings to specification. If, after reinstalling the unit, the discrepancy in oxygen readings persists, the problem is most likely elsewhere (i.e., flow blockage, primary device error, etc.). Further investigation should be made until the discrepancy in readings is resolved. The troubleshooting section of this manual may provide additional assistance in locating the problem.

Note: The MSDS on this material is available upon request through the Teledyne Environmental, Health and Safety Coordinator. Contact at (626) 934-1592

3.6.5 Anesthetic Gases

3.6.5.1GASES THAT INDUCE READING ERROR

When using the R17MED sensor in the presence of anesthetic gases such as Halothane, the oxygen reading may fall (see Table below). The magnitude of this error will depend upon the level of oxygen and the duration of exposure.

The anesthetic agents listed in the following table (Halothane, Enflurane, Isoflurane, Sevoflurane, and Desflurane) were vaporized into a stream of 30%

Teledyne Analytical Instruments

34

Image 34
Contents AX300-I AX300-I How To Use This Manual About This ManualSafety Messages Table of Contents Index Teledyne Analytical Instruments List of Figures AX300 Front ViewList of Tables Introduction AX300 Front ViewFeatures Applicable StandardsDigit LCD display. For RS-232 reconfiguration, see Section OptionsTheory of Operation ApplicationsSensor Signal Processing Setup OperationSensor Installation or Replacement Installing the R17MED SensorMounting Sensor Cable Connection to AnalyzerBattery Installation V-Mount Adapter InstallationInstalling Batteries Calibration Calibration SOperation AX300-I Output 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output PortProcedure UsePortable Oxygen Analyzer Operation Humidity TemperatureGas Sampling PressureDiscrepancy in Readings Anesthetic GasesOperation AX300-I Oxygen Reading Error in a Mixture of Anesthetic Gas Cleaning Do’s and Don’tsPortable Oxygen Analyzer Operation Blank Overall Maintenance General Service InformationBattery Maintenance Sensor Maintenance CalibrationHumidity Discrepancy in Readings Oxygen Reading Error Troubleshooting Troubleshooting√ Sensor Concentration of O2 is Error Codes WatchdogOther Problems with the Instrument Specifications Return Authorization for ServiceAppendix Appendix AX300-I Spare Parts List Optional AccessoriesInstrument for which the parts are intended Blank Index Index