Teledyne AX300-I manual Other Problems with the Instrument

Page 40

 

Service Manual

 

 

AX300-I

 

 

 

 

 

 

 

 

 

 

 

 

6 (5)

 

6

A stuck key has been detected. The second

 

 

 

(There is no

digit example (5) on the display shows

 

 

 

which key is stuck:

 

 

 

indication of

 

 

 

 

 

 

 

 

 

which key is

 

 

 

 

 

 

stuck)

1

- Batt

 

 

 

 

 

 

 

 

2

- Cal

 

 

 

 

5

- Silence

 

 

 

 

8

- Key Lock

 

 

 

 

 

 

7(2)

 

7

A failure has occurred during the automatic

 

 

 

(There is no

calibration of the digital to analog converter

 

 

 

(DAC) circuit. The second digit example (2)

 

 

 

indication of

 

 

 

shows the type of digital to analog failure.

 

 

 

the type of

 

 

 

0 - Measurement

 

 

 

DAC failure)

 

 

 

 

1

- High Test

 

 

 

 

2

- Low Test

 

 

 

 

3

- Offset Cal

 

 

 

 

4

- Gain Cal

 

 

 

 

 

 

 

3.9Other Problems with the Instrument

Most other problems arise from either mechanical damage from the instrument falling from a bench or table, or electronic component failure. In these units, repair or troubleshooting the PCB or individual component on the board is not feasible. It requires specialized test equipment and probes not generally available to the public. Under most circumstances a replacement of the entire PC Board is recommended. The instrument must be returned to the factory for PCB installation.

Occasionally, depending on the environment of use, keys can become stuck or function erratically due to contamination. Use a mild non-abrasive cleaner solution to periodically clean the keypad and screen. An aerosol jet spray of the type commonly used to clean computer keyboards can be used to dislodge dirt and accumulations from the keypad.

The Error Code Table (Table 3-3) includes a description of certain fault codes, which are diagnostic of some common (usually electronic) problem with the instrument. Some of these codes refer to specific components on the PCB that are problematic or have failed. These codes are useful in reporting

Teledyne Analytical Instruments

40

Image 40
Contents AX300-I AX300-I How To Use This Manual About This ManualSafety Messages Table of Contents Index Teledyne Analytical Instruments List of Figures AX300 Front ViewList of Tables Introduction AX300 Front ViewFeatures Applicable StandardsDigit LCD display. For RS-232 reconfiguration, see Section OptionsTheory of Operation ApplicationsSensor Signal Processing Setup OperationSensor Installation or Replacement Installing the R17MED SensorMounting Sensor Cable Connection to AnalyzerBattery Installation V-Mount Adapter InstallationInstalling Batteries Calibration Calibration SOperation AX300-I Output 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output PortProcedure UsePortable Oxygen Analyzer Operation Temperature Gas SamplingHumidity PressureDiscrepancy in Readings Anesthetic GasesOperation AX300-I Oxygen Reading Error in a Mixture of Anesthetic Gas Cleaning Do’s and Don’tsPortable Oxygen Analyzer Operation Blank Overall Maintenance General Service InformationBattery Maintenance Sensor Maintenance CalibrationHumidity Discrepancy in Readings Oxygen Reading Error Troubleshooting Troubleshooting√ Sensor Concentration of O2 is Error Codes WatchdogOther Problems with the Instrument Specifications Return Authorization for ServiceAppendix Appendix AX300-I Spare Parts List Optional AccessoriesInstrument for which the parts are intended Blank Index Index