Teledyne AX300-I manual Discrepancy in Readings, Anesthetic Gases

Page 25

Portable Oxygen Analyzer

Operation

 

 

 

It is recommended that the unit be calibrated prior to each use or every 8 hours.

When the sensor is connected to a ventilator circuit, the alternating “breathing” pressure cycles generated by the ventilator will be sensed as an increase in the oxygen percentage (especially if the sensor is fast enough to sense the changes, as is the R17MED). In reality, the percentage of oxygen is not changing; it is the total pressure that is increasing producing a corresponding increase in the partial pressure of oxygen. A hundred centimeter of water pressure pulse will produce a

0.11atmosphere, or an 11% increase in the total and therefore partial pressure of oxygen. Assuming that the sensor is fast enough to track this pressure pulse, an un-pressurized reading of 50% oxygen will increase to 55.3% if the sensor is subjected to a pressure cycle of 100cm H2O. The reading will rise proportionally less for smaller pressures.

2.3.4 Discrepancy in Readings

The AX300-I instrument should be used to measure the oxygen concentration exiting another oxygen mixing device or life support system (i.e., a blender, incubator or anesthesia machine). The information obtained from the AX300-I should never be used to adjust a life-support system, but should only be used as an indication that the life support system or device may require service and/or calibration.

When a discrepancy in oxygen readings is detected, the oxygen analyzers readings should be verified by checking the AX300-I battery condition and calibration using 100% oxygen. If the analyzer can be calibrated, the unit can be assumed to be in good working order and capable of providing readings to specification. If, after reinstalling the unit, the discrepancy in oxygen readings persists, the problem is most likely elsewhere (i.e., flow blockage, primary device error, etc.). Further investigation should be made until the discrepancy in readings is resolved. The troubleshooting section of this manual may provide additional assistance in locating the problem.

Note: The MSDS on this material is available upon request through the Teledyne Environmental, Health and Safety Coordinator. Contact at (626) 934-1592

2.3.5 Anesthetic Gases

2.3.5.1GASES THAT INDUCE READING ERROR

When using the R17MED sensor in the presence of anesthetic gases such as Halothane, the oxygen reading may fall (see Table below). The

Teledyne Analytical Instruments

25

Image 25
Contents AX300-I AX300-I About This Manual How To Use This ManualSafety Messages Table of Contents Index Teledyne Analytical Instruments AX300 Front View List of FiguresList of Tables AX300 Front View IntroductionApplicable Standards FeaturesOptions Digit LCD display. For RS-232 reconfiguration, see SectionTheory of Operation ApplicationsSensor Signal Processing Operation SetupInstalling the R17MED Sensor Sensor Installation or ReplacementSensor Cable Connection to Analyzer MountingV-Mount Adapter Installation Battery InstallationInstalling Batteries Calibration S CalibrationOperation AX300-I 0-1 VDC or RS 232 Digital Output Port Output 0-1 VDC or RS232 Use ProcedurePortable Oxygen Analyzer Operation Gas Sampling TemperatureHumidity PressureAnesthetic Gases Discrepancy in ReadingsOperation AX300-I Oxygen Reading Error in a Mixture of Anesthetic Gas Do’s and Don’ts CleaningPortable Oxygen Analyzer Operation Blank Overall Maintenance General Service InformationBattery Maintenance Calibration Sensor MaintenanceHumidity Discrepancy in Readings Oxygen Reading Error Troubleshooting Troubleshooting√ Sensor Concentration of O2 is Watchdog Error CodesOther Problems with the Instrument Specifications Return Authorization for ServiceAppendix Appendix AX300-I Optional Accessories Spare Parts ListInstrument for which the parts are intended Blank Index Index