Service Manual |
| |
|
|
|
3.6.5.2CARE AFTER USE IN NITROUS OXIDE
CAUTION: THE R17MED SENSOR SHOULD NOT BE LEFT IN NITROUS OXIDE MIXTURES ANY LONGER THAN ABSOLUTELY NECESSARY.
After exposure to nitrous oxide mixtures, the sensor should be left in 100% oxygen overnight (e.g., left in a breathing circuit that has been flushed with pure oxygen). If this is not practical, when using the tee, remove the plastic flow diverter and leave the sensor in room air. If the oxygen reading continues to drop after each use in nitrous oxide the sensor should be removed from service. If the sensor can no longer be calibrated or if there is any sign of electrolyte leakage, the sensor should be disposed of in accordance with local regulations and the Material Safety Data Sheet (MSDS).
Note: The MSDS on this material is available upon request through the Teledyne Environmental, Health and Safety Coordinator. Contact at (626)
3.7 Troubleshooting
The
Table 3-2 Troubleshooting
Symptom | Why | What To Do |
New sensor responds | If the sensor is new | A) Wait |
slow or drifts. | and was just | recalibrate. |
| removed from its |
|
| sealed bag it may |
|
| need to run for |
|
| several hours. |
|
Sensor will not read | Calibration in 100% | A) Recalibrate using dry gas |
below 22 % after | was invalid or the | making sure the reading |
calibration in 100% O2. | room air is | stabilizes before making |
| contaminated with | any adjustments. |
| excess oxygen. | B) Make sure that at least 6" (30 |
|
| |
|
| cm.) of tubing is attached to |
|
| the exhaust side of the tee |
Teledyne Analytical Instruments | 36 |