Teledyne AX300-I manual Portable Oxygen Analyzer Operation

Page 23

Portable Oxygen Analyzer

Operation

 

 

 

BACKPRESSURE OR RESTRICTION TO FLOW.

ERRORS IN READINGS WILL RESULT IF THIS IS

NOT FOLLOWED.

THE OXYGEN SENSOR IS A NON-STERILE DEVICE AND SHOULD BE USED IN CONJUCTION WITH AN ANTIBACTERIAL FILTER. ALWAYS INSTALL THE SENSOR ON THE INSPIRED LINE AHEAD OF FILTERS, HUMIDIFIERS AND MEDICATING DEVICES. NEVER INSTALL THE SENSOR IN A LOCATION THAT WILL EXPOSE THE SENSOR TO PATIENTS EXHALED BREATH OR SECRETIONS UNLESS YOU INTEND TO DISPOSE OF THE SENSOR AND FLOW ADAPTER AFTER USE.

When analyzing for oxygen in confined volumes such as incubators, hoods, etc., the flow diverter must be removed from the R17MED sensor so that it does not interfere with the rapid exchange of gases to and from the sensing surface of the sensor.

CAUTION: FAILURE TO REMOVE THE DIVERTER IN THESE APPLICATION AREAS WILL RESULT IN A MARKED LOWERING OF THE RESPONSE TIME OF THE SENSOR.

The R17MED sensor can be placed or hung inside incubators, tents, etc. When it is necessary to thread the cable through a small hole in order to gain access to the inside of a chamber, the cable should be disconnected at the sensor, threaded through the hole, and reconnected inside the chamber.

The LOCK/UNLOCK key can be used to lock out any accidental interference to the front panel keys. The LOCK/UNLOCK key acts as a toggle, pressing LOCK/UNLOCK once renders inactive all keys except the ALARM SILENCE () key. Pressing LOCK/UNLOCK a second time unlocks the keypad.

Teledyne Analytical Instruments

23

Image 23
Contents AX300-I AX300-I About This Manual How To Use This ManualSafety Messages Table of Contents Index Teledyne Analytical Instruments AX300 Front View List of FiguresList of Tables AX300 Front View IntroductionApplicable Standards FeaturesOptions Digit LCD display. For RS-232 reconfiguration, see SectionSensor ApplicationsTheory of Operation Signal Processing Operation SetupInstalling the R17MED Sensor Sensor Installation or ReplacementSensor Cable Connection to Analyzer MountingV-Mount Adapter Installation Battery InstallationInstalling Batteries Calibration S CalibrationOperation AX300-I 0-1 VDC or RS 232 Digital Output Port Output 0-1 VDC or RS232Use ProcedurePortable Oxygen Analyzer Operation Pressure TemperatureGas Sampling HumidityAnesthetic Gases Discrepancy in ReadingsOperation AX300-I Oxygen Reading Error in a Mixture of Anesthetic Gas Do’s and Don’ts CleaningPortable Oxygen Analyzer Operation Blank Battery Maintenance General Service InformationOverall Maintenance Calibration Sensor MaintenanceHumidity Discrepancy in Readings Oxygen Reading Error Troubleshooting Troubleshooting√ Sensor Concentration of O2 is Watchdog Error CodesOther Problems with the Instrument Appendix Return Authorization for ServiceSpecifications Appendix AX300-I Optional Accessories Spare Parts ListInstrument for which the parts are intended Blank Index Index