Teledyne AX300-I manual Watchdog, Error Codes

Page 39

Portable Oxygen Analyzer

Service Manual

 

 

 

Note: In the event that none of these procedures produce desired results, remove the batteries and return the unit to Teledyne for repair.

3.8 Watchdog

The AX300-I is equipped with a watchdog circuit that continuously analyzers the electronics for proper operation. If the watchdog detects a failure, one of the following codes will appear on the LCD.

The error codes can appear on the LCD when batteries are first installed, during normal operation or if the unit is subjected to extreme shock. In some cases an additional digit is used in the error code to supply additional information. For example, the error code 6 and 7 are followed by another digit listed as (N) in Table 3.3. The error code 65 would indicate that a key is stuck and this key is the Silence key.

In addition to supplying visual error codes, the audio device will beep a number of times to indicate the general error in case the display is not functional.

Note: To reset the watchdog error code. Remove one battery for 5 seconds and replace. If the error persists contact your locale representative or Teledyne for assistance.

Table 3-3 Error Codes

 

Error Code

Audio Beeps

Error

 

Indeterminate

2

The watchdog timer has timed out indicating

 

 

 

a serious software error

 

30

3

Analog output is different from the expected

 

 

 

value. May indicate a shorted or over-loaded

 

 

 

analog output or a failure of the analog to

 

 

 

digital converter or digital to analog

 

 

 

converter circuit.

 

50

5

The ADC circuit failed during POST.

 

 

 

 

Teledyne Analytical Instruments

39

Image 39
Contents AX300-I AX300-I About This Manual How To Use This ManualSafety Messages Table of Contents Index Teledyne Analytical Instruments AX300 Front View List of FiguresList of Tables AX300 Front View IntroductionApplicable Standards FeaturesOptions Digit LCD display. For RS-232 reconfiguration, see SectionApplications Theory of OperationSensor Signal Processing Operation SetupInstalling the R17MED Sensor Sensor Installation or ReplacementSensor Cable Connection to Analyzer MountingV-Mount Adapter Installation Battery InstallationInstalling Batteries Calibration S CalibrationOperation AX300-I 0-1 VDC or RS 232 Digital Output Port Output 0-1 VDC or RS232Use ProcedurePortable Oxygen Analyzer Operation Pressure TemperatureGas Sampling HumidityAnesthetic Gases Discrepancy in ReadingsOperation AX300-I Oxygen Reading Error in a Mixture of Anesthetic Gas Do’s and Don’ts CleaningPortable Oxygen Analyzer Operation Blank General Service Information Overall MaintenanceBattery Maintenance Calibration Sensor MaintenanceHumidity Discrepancy in Readings Oxygen Reading Error Troubleshooting Troubleshooting√ Sensor Concentration of O2 is Watchdog Error CodesOther Problems with the Instrument Return Authorization for Service SpecificationsAppendix Appendix AX300-I Optional Accessories Spare Parts ListInstrument for which the parts are intended Blank Index Index