Teledyne AX300-I manual Operation, Setup

Page 14

Operation

AX300-I

 

 

 

Operation

Note: Upon receipt, INSPECT THE ENTIRE UNIT FOR DAMAGE. Check the unit and all included accessories for broken or loose parts. If damaged, DO NOT USE. Notify the shipper, and consult Teledyne Analytical Instruments.

Note: This equipment is internally powered using 3 AA batteries.

CAUTION: THE AX300-I, OXYGEN SENSOR AND ASSOCIATED HARDWARE ARE NON-STERILE DEVICES. DO NOT AUTOCLAVE THE INSTRUMENT OR SENSOR, AS THIS WILL DAMAGE THE EQUIPMENT.

2.1 Setup

The AX300-I Portable Oxygen Analyzer is suitable for use in many medical applications. The unit is equipped with a stand and can be used on a tabletop or wall mounted using mounting bracket. An optional pole mount clamp is also available.

To set up and use your AX300-I analyzer:

1.Install the sensor.

2.Install the batteries.

3.Calibrate the unit.

The control keys are designed for easy operation. A LOCK/UNLOCK key has been supplied to prevent accidental changes to critical settings. This eliminates unwanted changes in calibration or alarm settings from accidental touching or bumping of the keys. To further reduce the possibly of incorrect adjustments at least two keys must be pressed in order to modify a critical calibration set point value.

Note: The ALARM SILENCE () key continues to operate normally when the lock feature is activated.

Teledyne Analytical Instruments

14

Image 14
Contents AX300-I AX300-I How To Use This Manual About This ManualSafety Messages Table of Contents Index Teledyne Analytical Instruments List of Figures AX300 Front ViewList of Tables Introduction AX300 Front ViewFeatures Applicable StandardsDigit LCD display. For RS-232 reconfiguration, see Section OptionsSensor ApplicationsTheory of Operation Signal Processing Setup OperationSensor Installation or Replacement Installing the R17MED SensorMounting Sensor Cable Connection to AnalyzerBattery Installation V-Mount Adapter InstallationInstalling Batteries Calibration Calibration SOperation AX300-I Output 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output PortProcedure UsePortable Oxygen Analyzer Operation Humidity TemperatureGas Sampling PressureDiscrepancy in Readings Anesthetic GasesOperation AX300-I Oxygen Reading Error in a Mixture of Anesthetic Gas Cleaning Do’s and Don’tsPortable Oxygen Analyzer Operation Blank Battery Maintenance General Service InformationOverall Maintenance Sensor Maintenance CalibrationHumidity Discrepancy in Readings Oxygen Reading Error Troubleshooting Troubleshooting√ Sensor Concentration of O2 is Error Codes WatchdogOther Problems with the Instrument Appendix Return Authorization for ServiceSpecifications Appendix AX300-I Spare Parts List Optional AccessoriesInstrument for which the parts are intended Blank Index Index