Teledyne AX300-I manual Oxygen Reading Error

Page 35

Portable Oxygen Analyzer

Service Manual

 

 

 

oxygen / 70% nitrous oxide, and the resulting drops in oxygen level after an exposure of approximately two hours were noted.

Exposures in excess of two hours may produce slightly greater errors. The errors listed are typical for all oxygen sensors such as the R17MED. Exposing the sensor to air or gases that do not contain anesthetic agents for a period of time equal to or greater than the exposure interval will eliminate the reading error in most cases.

Table 3-1: Oxygen Reading Error in a Mixture of Anesthetic Gas

Gas or Vapor Level

(Balance: Mixture of 30% O2 / 70% N2O, except where noted)

Gas or Vapor

Test Level

 

 

Helium

50%, balance O2

Nitrous Oxide

80%, balance O2

Carbon Dioxide

10%, balance O2

Halothane

4%

Enflurane

5%

Isoflurane

5%

Sevoflurane

5%

Desflurane

15%

 

 

Oxygen Reading

Error

0%

0%

0%

<1.5% O2 *

<1.5% O2 *

<1.5% O2 *

<1.5% O2 *

<1.5% O2 *

*Errors are approximate and may vary based on exposure times and concentrations.

These performances meet or exceed the requirements of ISO 7767: 1997 (E).

CAUTION: THE AX300-I SHOULD NOT BE USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS SUCH AS DIETHYL ETHER OR CYCLOPROPANE.

CAUTION: THE AX300-I, OXYGEN SENSOR AND ASSOCIATED HARDWARE ARE NON-STERILE DEVICES. DO NOT AUTOCLAVE THE INSTRUMENT OR SENSOR, AS THIS WILL DAMAGE THE EQUIPMENT.

Teledyne Analytical Instruments

35

Image 35
Contents AX300-I AX300-I About This Manual How To Use This ManualSafety Messages Table of Contents Index Teledyne Analytical Instruments AX300 Front View List of FiguresList of Tables AX300 Front View IntroductionApplicable Standards FeaturesOptions Digit LCD display. For RS-232 reconfiguration, see SectionSensor ApplicationsTheory of Operation Signal Processing Operation SetupInstalling the R17MED Sensor Sensor Installation or ReplacementSensor Cable Connection to Analyzer MountingV-Mount Adapter Installation Battery InstallationInstalling Batteries Calibration S CalibrationOperation AX300-I 0-1 VDC or RS 232 Digital Output Port Output 0-1 VDC or RS232Use ProcedurePortable Oxygen Analyzer Operation Pressure TemperatureGas Sampling HumidityAnesthetic Gases Discrepancy in ReadingsOperation AX300-I Oxygen Reading Error in a Mixture of Anesthetic Gas Do’s and Don’ts CleaningPortable Oxygen Analyzer Operation Blank Battery Maintenance General Service InformationOverall Maintenance Calibration Sensor MaintenanceHumidity Discrepancy in Readings Oxygen Reading Error Troubleshooting Troubleshooting√ Sensor Concentration of O2 is Watchdog Error CodesOther Problems with the Instrument Appendix Return Authorization for ServiceSpecifications Appendix AX300-I Optional Accessories Spare Parts ListInstrument for which the parts are intended Blank Index Index