Teledyne AX300-I manual Installing Batteries

Page 18

Operation

AX300-I

 

 

 

1.Turn the unit off (if it is on).

2.Hold the instrument face down in the palm of your hand. Use a coin to pry up the latch that secures the battery compartment door. Remove the battery compartment door.

CAUTION: IMPROPER INSTALLATION OF THE BATTERIES MAY RESULT IN DAMAGE TO THE UNIT AND BATTERIES.

Note: Use alkaline batteries only. Other battery types will produce erroneous battery test readings.

3.Install 3 “AA” alkaline batteries into the holder as shown in Figure 2-6. Each battery has its own slot. To insure proper polarity, place the bottom (flat) or negative end of the battery in the end of the holder marked “–”. Place the top (button) or positive end of the battery in the end of the holder marked “+”. Do this for each battery.

Figure 2-6: Installing Batteries

4.Re-install the battery compartment door. When the unit is first turned on the display will momentarily display all LCD segments. During this period diagnostic tests are being conducted to insure the circuits are functioning correctly. The unit will activate the audible and visual alarms for about

1 second. The LCD will flash continuously indicating the unit is in the unlocked position ready for calibration.

Note: When batteries are first installed or power is lost for any reason the instrument defaults to the calibration mode. All

Teledyne Analytical Instruments

18

Image 18
Contents AX300-I AX300-I How To Use This Manual About This ManualSafety Messages Table of Contents Index Teledyne Analytical Instruments List of Figures AX300 Front ViewList of Tables Introduction AX300 Front ViewFeatures Applicable StandardsDigit LCD display. For RS-232 reconfiguration, see Section OptionsApplications Theory of OperationSensor Signal Processing Setup OperationSensor Installation or Replacement Installing the R17MED SensorMounting Sensor Cable Connection to AnalyzerBattery Installation V-Mount Adapter InstallationInstalling Batteries Calibration Calibration SOperation AX300-I Output 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output PortProcedure UsePortable Oxygen Analyzer Operation Humidity TemperatureGas Sampling PressureDiscrepancy in Readings Anesthetic GasesOperation AX300-I Oxygen Reading Error in a Mixture of Anesthetic Gas Cleaning Do’s and Don’tsPortable Oxygen Analyzer Operation Blank General Service Information Overall MaintenanceBattery Maintenance Sensor Maintenance CalibrationHumidity Discrepancy in Readings Oxygen Reading Error Troubleshooting Troubleshooting√ Sensor Concentration of O2 is Error Codes WatchdogOther Problems with the Instrument Return Authorization for Service SpecificationsAppendix Appendix AX300-I Spare Parts List Optional AccessoriesInstrument for which the parts are intended Blank Index Index