Portable Oxygen Analyzer | Operation | |
|
|
|
Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas
Gas or Vapor Level
(Balance: Mixture of 30% O2 / 70% N2O, except where noted)
| Gas or Vapor | Test Level | Oxygen Reading |
|
|
|
| Error |
|
| Helium | 50%, balance O2 | 0% |
|
| Nitrous Oxide | 80%, balance O2 | 0% |
|
| Carbon Dioxide | 10%, balance O2 | 0% |
|
| Halothane | 4% | < 1.5% O2 * |
|
| Enflurane | 5% | < 1.5% O2 * |
|
| Isoflurane | 5% | < 1.5% O2 * |
|
| Sevoflurane | 5% | < 1.5% O2 * |
|
| Desflurane | 15% | < 1.5% O2 * |
|
*Errors are approximate and may vary based on exposure times and concentrations.
These performances meet or exceed the requirements of ISO 7767: 1997 (E).
CAUTION: THE
CAUTION: THE
2.3.5.2CARE AFTER USE IN NITROUS OXIDE
CAUTION: THE R17MED SENSOR SHOULD NOT BE LEFT IN NITROUS OXIDE MIXTURES ANY LONGER THAN ABSOLUTELY NECESSARY.
After exposure to nitrous oxide mixtures, the sensor should be left in 100% oxygen overnight (e.g., left in a breathing circuit that has been flushed with pure oxygen). If this is not practical, when using the tee, remove the plastic flow diverter and leave the sensor in room air. If the
Teledyne Analytical Instruments | 27 |