Assembling the VPAP Adapt SV System

Complete the following steps to assemble the VPAP Adapt SV system. Make sure the area around the unit is clean (dust free) and clear of bedding, clothes and any other potential blockages.

To add a humidifier see page 11.

!

1

WARNING

Use only the air tubing system supplied with the VPAP Adapt SV system.

Connect the power cord

There are two sockets at the rear of the flow generator—the top one for an AC (standard mains electricity) and the lower one for a DC power cord (see “Using DC Power to Run the VPAP Adapt SV” on page 16). ResMed recommends using the AC power cord supplied with the unit.

The customized ResMed power cord can be held in place with the locking clips. To insert the locking clip, pinch the free ends together and fit the pins into the holes on both sides of the socket. Insert the power cord into the socket. Push the locking clip down so that the groove holds the power cord in place. Plug one end of the power cord into the power socket at the rear of the unit, and the other end into a power outlet.

Power switch

AC locking clip

AC power cord

Plug the free end of the power cord into a power outlet.

CAUTIONS

! Do not connect both AC and DC power cords to the VPAP Adapt SV at the same time.

Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord.

WARNING

! Make sure the power cord and plug are in good condition and the equipment is not damaged.

VPAP Adapt SV System

9

Image 15
Contents Vpap Adapt SV USA ENG Reorder number 26869/2 06 Page Contents Daily Weekly Periodically Air Filter Definitions IntroductionVpap Adapt SV User/Owner ResponsibilityPage Page Page Quick Setup Guide Page Vpap Adapt SV Components Vpap Adapt SV SystemAir tubing fully assembled MasksAssembling the Vpap Adapt SV System Connect the power cordAttach the mask to the air tubing Humidifiers Connecting a HumidifierHumidAire 2i attached to the Vpap Adapt SV HumidAireHumidAire Page Anti-bacterial Filter ResLinkSupplemental Oxygen Using DC Power to Run the Vpap Adapt SVStopping Treatment Using the Vpap Adapt SVStandby Mode Starting TreatmentUp/Down key Control Panel FunctionsLCD screen Start/Stop keyRight key red Adjusting SettingsAlarm LEDs and Alarm Mute Left key greenExample You wish to change the setting for the mask you are usingMenus Navigating the Patient menu Using the Vpap Adapt SVSettings Menu MaskLearn Circuit Leak Alert Ramp Cpap mode onlyResults Menu SmartStopOptions Menu Servicing Menu Treatment ScreensPage Triggered by AlarmsLow Mask Pressure Alarm If a low pressure alarm soundsPower Fail Alarm High Pressure AlarmHigh Leak Alarm If a flow blocked alarm sounds Insufficient Pressure Support AlarmFlow Blocked Alarm If an insufficient pressure support alarm soundsPage Cleaning and Maintenance DailyWeekly Periodically Air tubing assembly removed for cleaningFitting the filter Air FilterServicing Hypoallergenic Air FilterPage Unusual manner TroubleshootingPage Glossary AlarmsPage Technical Specifications IEC 601-1 classifications Inspiratory Trigger CharacteristicsPage Glossary of symbols Mask setting flow/pressure characteristicsEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index