ResMed Adapt SV System manual Separation distance according to frequency of transmitter

Page 52

Recommended separation distances between portable and mobile RF communications equipment and the VPAP Adapt SV

The VPAP Adapt SV is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the VPAP Adapt SV can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the VPAP Adapt SV as recommended below, according to the maximum output power of the communications equipment.

 

Separation distance according to frequency of transmitter

 

m

 

 

 

 

 

 

Rated maximum output

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

power of transmitter

d = 1.17 P

d = 0.35 P

d = 0.70 P

W

 

 

 

 

0.01

0.17

0.04

0.07

 

 

 

 

0.1

0.37

0.11

0.22

 

 

 

 

1

1.17

0.35

0.70

 

 

 

 

10

3.69

1.11

2.21

 

 

 

 

100

11.70

3.50

7.00

 

 

 

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

46

Image 52
Contents Vpap Adapt SV USA ENG Reorder number 26869/2 06 Page Contents Daily Weekly Periodically Air Filter Introduction Vpap Adapt SVUser/Owner Responsibility DefinitionsPage Page Page Quick Setup Guide Page Vpap Adapt SV System Vpap Adapt SV ComponentsMasks Air tubing fully assembledConnect the power cord Assembling the Vpap Adapt SV SystemAttach the mask to the air tubing Connecting a Humidifier HumidifiersHumidAire HumidAire 2i attached to the Vpap Adapt SVHumidAire Page ResLink Anti-bacterial FilterUsing DC Power to Run the Vpap Adapt SV Supplemental OxygenUsing the Vpap Adapt SV Standby ModeStarting Treatment Stopping TreatmentControl Panel Functions LCD screenStart/Stop key Up/Down keyAdjusting Settings Alarm LEDs and Alarm MuteLeft key green Right key redYou wish to change the setting for the mask you are using ExampleNavigating the Patient menu Using the Vpap Adapt SV MenusMask Settings MenuLearn Circuit Ramp Cpap mode only Results MenuSmartStop Leak AlertOptions Menu Treatment Screens Servicing MenuPage Alarms Low Mask Pressure AlarmIf a low pressure alarm sounds Triggered byHigh Pressure Alarm Power Fail AlarmHigh Leak Alarm Insufficient Pressure Support Alarm Flow Blocked AlarmIf an insufficient pressure support alarm sounds If a flow blocked alarm soundsPage Daily Cleaning and MaintenanceWeekly Air tubing assembly removed for cleaning PeriodicallyAir Filter ServicingHypoallergenic Air Filter Fitting the filterPage Troubleshooting Unusual mannerPage Alarms GlossaryPage Technical Specifications Inspiratory Trigger Characteristics IEC 601-1 classificationsPage Mask setting flow/pressure characteristics Glossary of symbolsEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index