impaired ability to clear secretions

pneumothorax or pneumomediastinum

recent cranial trauma or surgery

chronic hypoventilation.

Below are general warnings and cautions. Further specific warnings, cautions and notes appear next to the relevant instructions in the manual.

!Warnings

The VPAP Adapt SV is NOT a life support ventilator.

The entire manual should be read before using the VPAP Adapt SV.

Advice contained in this manual should not supersede instructions given by the prescribing physician.

The VPAP Adapt SV should be used with masks and accessories recommended by ResMed or the prescribing physician. Use of incorrect

masks and accessories may adversely affect the function of the

VPAP Adapt SV.

The VPAP Adapt SV is designed for use with masks that allow exhaled gases to be flushed out through vent holes. Exhaled gases will be rebreathed if the mask is worn with the machine turned off, or the vent holes are occluded. If this occurs over prolonged periods, suffocation may occur.

In the event of power failure or machine malfunction, remove the mask.

The air flow for breathing produced by this device can be as much as 11oF (6oC ) higher than the temperature of the room. Caution should be exercised if the room temperature is warmer than 90oF (32oC).

The VPAP Adapt SV can be set to deliver pressures up to 20 cm H2O. In the unlikely event of certain fault conditions, pressures of up to 40 cm H2O for less than one second are possible.

The VPAP Adapt SV is not suitable for use in the vicinity of flammable anesthetics.

If oxygen is used with the VPAP Adapt SV, the oxygen flow should be stopped when the device is not operating. If oxygen flow continues when the device is not operating, oxygen may accumulate within the device and create a risk of fire.

Do not use the VPAP Adapt SV if there are obvious external defects or unexplained changes in performance.

Do not open the VPAP Adapt SV case. There are no user serviceable parts inside. Repairs and internal servicing should only be performed by an authorized service agent.

!Cautions

You should report unusual chest pain, severe headache or increased breathlessness to your physician. An acute upper respiratory tract infection may require temporary discontinuation of treatment.

2

Image 8
Contents Vpap Adapt SV USA ENG Reorder number 26869/2 06 Page Contents Daily Weekly Periodically Air Filter Introduction Vpap Adapt SVUser/Owner Responsibility DefinitionsPage Page Page Quick Setup Guide Page Vpap Adapt SV System Vpap Adapt SV ComponentsMasks Air tubing fully assembledConnect the power cord Assembling the Vpap Adapt SV SystemAttach the mask to the air tubing Connecting a Humidifier HumidifiersHumidAire HumidAire 2i attached to the Vpap Adapt SVHumidAire Page ResLink Anti-bacterial FilterUsing DC Power to Run the Vpap Adapt SV Supplemental OxygenUsing the Vpap Adapt SV Standby ModeStarting Treatment Stopping TreatmentControl Panel Functions LCD screenStart/Stop key Up/Down keyAdjusting Settings Alarm LEDs and Alarm MuteLeft key green Right key redYou wish to change the setting for the mask you are using ExampleNavigating the Patient menu Using the Vpap Adapt SV MenusLearn Circuit Settings MenuMask Ramp Cpap mode only Results MenuSmartStop Leak AlertOptions Menu Treatment Screens Servicing MenuPage Alarms Low Mask Pressure AlarmIf a low pressure alarm sounds Triggered byHigh Leak Alarm Power Fail AlarmHigh Pressure Alarm Insufficient Pressure Support Alarm Flow Blocked AlarmIf an insufficient pressure support alarm sounds If a flow blocked alarm soundsPage Weekly Cleaning and MaintenanceDaily Air tubing assembly removed for cleaning PeriodicallyAir Filter ServicingHypoallergenic Air Filter Fitting the filterPage Troubleshooting Unusual mannerPage Alarms GlossaryPage Technical Specifications Inspiratory Trigger Characteristics IEC 601-1 classificationsPage Mask setting flow/pressure characteristics Glossary of symbolsEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index