ResMed Adapt SV System manual Technical Specifications

Page 45

Technical Specifications

Dimensions (H x W x D)

5.6" x 9.8" x 11.5" (142 mm x 248 mm x 293 mm)

Weight

Flow generator with cover: 8.14 lb (3.7 kg)

Flow generator with Humidaire 2i (empty): 9.24 lb (4.2 kg)

Environmental conditions

Operating temperature: 41oF to 104oF (+5oC to +40oC)

Storage and transport temperature: -4oF to +140oF (-20oC to +60oC)

Humidity—operating, storage and transportation: 10%–95% non-condensing

Housing construction

Injection moulded plastic

Air tubing

1 x 6’6” (2 m)

Air filter

Synthetic fiber

Patient connection port

22 mm taper, compatible with EN 1281-1:1997 Anaesthetic & Respiratory Equipment —Conical Connectors

Power supply

AC Input range: 110–120 V and 220–240 V; 50–60 Hz; 60 VA

DC Input range (if using DC converter): 12–13.8 V DC

Performance

Operating pressure range: 4–20 cm H2O

Sound pressure level

<30 dBA (tested in accordance with the requirements of ISO 17510-1)

Battery type

Lithium

Alarm specifications

High maximum mask pressure alarm Stops air delivery at 25 cm H2O >700 ms.

High mean mask pressure alarm

Stops air delivery when mean mask pressure exceeds 15 cm H2O (averaged over approximately 1 minute).

Low mask pressure alarm

Mask pressure is less than 3 cm H2O for longer than 5 seconds while the patient is using the VPAP Adapt SV.

Technical Specifications

39

Image 45
Contents Vpap Adapt SV USA ENG Reorder number 26869/2 06 Page Contents Daily Weekly Periodically Air Filter Vpap Adapt SV IntroductionUser/Owner Responsibility DefinitionsPage Page Page Quick Setup Guide Page Vpap Adapt SV Components Vpap Adapt SV SystemAir tubing fully assembled MasksAssembling the Vpap Adapt SV System Connect the power cordAttach the mask to the air tubing Humidifiers Connecting a HumidifierHumidAire 2i attached to the Vpap Adapt SV HumidAireHumidAire Page Anti-bacterial Filter ResLinkSupplemental Oxygen Using DC Power to Run the Vpap Adapt SVStandby Mode Using the Vpap Adapt SVStarting Treatment Stopping TreatmentLCD screen Control Panel FunctionsStart/Stop key Up/Down keyAlarm LEDs and Alarm Mute Adjusting SettingsLeft key green Right key redExample You wish to change the setting for the mask you are usingMenus Navigating the Patient menu Using the Vpap Adapt SVSettings Menu MaskLearn Circuit Results Menu Ramp Cpap mode onlySmartStop Leak AlertOptions Menu Servicing Menu Treatment ScreensPage Low Mask Pressure Alarm AlarmsIf a low pressure alarm sounds Triggered byPower Fail Alarm High Pressure AlarmHigh Leak Alarm Flow Blocked Alarm Insufficient Pressure Support AlarmIf an insufficient pressure support alarm sounds If a flow blocked alarm soundsPage Cleaning and Maintenance DailyWeekly Periodically Air tubing assembly removed for cleaningServicing Air FilterHypoallergenic Air Filter Fitting the filterPage Unusual manner TroubleshootingPage Glossary AlarmsPage Technical Specifications IEC 601-1 classifications Inspiratory Trigger CharacteristicsPage Glossary of symbols Mask setting flow/pressure characteristicsEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index