Technical Specifications
Dimensions (H x W x D)
5.6" x 9.8" x 11.5" (142 mm x 248 mm x 293 mm)
Weight
Flow generator with cover: 8.14 lb (3.7 kg)
Flow generator with Humidaire 2i (empty): 9.24 lb (4.2 kg)
Environmental conditions
Operating temperature: 41oF to 104oF (+5oC to +40oC)
Storage and transport temperature:
Housing construction
Injection moulded plastic
Air tubing
1 x 6’6” (2 m)
Air filter
Synthetic fiber
Patient connection port
22 mm taper, compatible with EN
Power supply
AC Input range:
DC Input range (if using DC converter):
Performance
Operating pressure range:
Sound pressure level
<30 dBA (tested in accordance with the requirements of ISO
Battery type
Lithium
Alarm specifications
•High maximum mask pressure alarm Stops air delivery at 25 cm H2O >700 ms.
•High mean mask pressure alarm
Stops air delivery when mean mask pressure exceeds 15 cm H2O (averaged over approximately 1 minute).
•Low mask pressure alarm
Mask pressure is less than 3 cm H2O for longer than 5 seconds while the patient is using the VPAP Adapt SV.
Technical Specifications | 39 |