ResMed Adapt SV System manual ResLink, Anti-bacterial Filter

Page 21

ResLink™

The ResLink is a device that records data onto a SmartMedia™ card during your

therapy. The card can be returned to your clinician so that they can observe your progress and adjust flow generator parameters if necessary. The ResLink connects to the back of VPAP Adapt SV as shown. For full details on using the ResLink, please refer to the ResLink User’s Manual.

Anti-bacterial Filter

Your clinician may recommend that you use an anti-bacterial filter. A filter should be placed in the air tubing between the mask and the flow generator. If using a humidifier, place the filter between it and the flow generator (see Figure 10 below).

Typically, the filter should be replaced daily. Check the filter instructions for full details.

Whenever a filter is added, removed or replaced you must run the Air Circuit Learn function (see “Learn Circuit” on page 22).

Note: Only hydrophobic anti-bacterial filters should be used with a HumidAire 2i.

Anti-bacterial filters

Figure 10: Inserting anti-bacterial filters

VPAP Adapt SV System

15

Image 21
Contents Vpap Adapt SV USA ENG Reorder number 26869/2 06 Page Contents Daily Weekly Periodically Air Filter Vpap Adapt SV IntroductionUser/Owner Responsibility DefinitionsPage Page Page Quick Setup Guide Page Vpap Adapt SV Components Vpap Adapt SV SystemAir tubing fully assembled MasksAssembling the Vpap Adapt SV System Connect the power cordAttach the mask to the air tubing Humidifiers Connecting a Humidifier HumidAire 2i attached to the Vpap Adapt SV HumidAireHumidAire Page Anti-bacterial Filter ResLinkSupplemental Oxygen Using DC Power to Run the Vpap Adapt SVStandby Mode Using the Vpap Adapt SVStarting Treatment Stopping TreatmentLCD screen Control Panel FunctionsStart/Stop key Up/Down keyAlarm LEDs and Alarm Mute Adjusting SettingsLeft key green Right key redExample You wish to change the setting for the mask you are usingMenus Navigating the Patient menu Using the Vpap Adapt SVSettings Menu MaskLearn Circuit Results Menu Ramp Cpap mode onlySmartStop Leak AlertOptions Menu Servicing Menu Treatment ScreensPage Low Mask Pressure Alarm AlarmsIf a low pressure alarm sounds Triggered byPower Fail Alarm High Pressure AlarmHigh Leak Alarm Flow Blocked Alarm Insufficient Pressure Support AlarmIf an insufficient pressure support alarm sounds If a flow blocked alarm soundsPage Cleaning and Maintenance DailyWeekly Periodically Air tubing assembly removed for cleaningServicing Air FilterHypoallergenic Air Filter Fitting the filterPage Unusual manner TroubleshootingPage Glossary AlarmsPage Technical Specifications IEC 601-1 classifications Inspiratory Trigger CharacteristicsPage Glossary of symbols Mask setting flow/pressure characteristicsEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index