ResMed Adapt SV System manual Insufficient Pressure Support Alarm, Flow Blocked Alarm

Page 35

check that the air tube and sensor tube are connected properly.

Will stop when:

the mask leak has been rectified.

Note: The High Leak alarm can be turned off in the Leak Alert option in the Settings menu.

Insufficient Pressure Support Alarm

Triggered by:

air pressure at the mask has failed to reach an expected level for 3 breaths

the Air Circuit Learn function (page 22) has not been run after adding a new component (eg, new mask, humidifier) to the system.

If an insufficient pressure support alarm sounds:

check that the pressure sensor tubing is not kinked

check that the pressure sensor tubing is properly connected

run LEARN CIRCUIT.

Will stop when:

the pressure sensor tubing is unobstructed and properly connected to the flow generator and mask

treatment is stopped (Start/Stop key or SmartStop)

the VPAP Adapt SV is turned off.

Flow Blocked Alarm

Triggered by:

blockage in air circuit.

If a flow blocked alarm sounds:

1.The treatment will stop.

2.Turn power off.

3.Check whether there is a blockage in the air circuit.

4.Remove blockage.

5.Turn power back on.

6.If the flow blocked alarm activates repeatedly, discontinue use and return to ResMed for servicing. If the alarm does not recur, then continue to use as normal.

Will stop when:

blockage is removed.

Alarms 29

Image 35
Contents Vpap Adapt SV USA ENG Reorder number 26869/2 06 Page Contents Daily Weekly Periodically Air Filter Definitions IntroductionVpap Adapt SV User/Owner ResponsibilityPage Page Page Quick Setup Guide Page Vpap Adapt SV Components Vpap Adapt SV SystemAir tubing fully assembled MasksAssembling the Vpap Adapt SV System Connect the power cordAttach the mask to the air tubing Humidifiers Connecting a HumidifierHumidAire 2i attached to the Vpap Adapt SV HumidAireHumidAire Page Anti-bacterial Filter ResLinkSupplemental Oxygen Using DC Power to Run the Vpap Adapt SVStopping Treatment Using the Vpap Adapt SVStandby Mode Starting TreatmentUp/Down key Control Panel FunctionsLCD screen Start/Stop keyRight key red Adjusting SettingsAlarm LEDs and Alarm Mute Left key greenExample You wish to change the setting for the mask you are usingMenus Navigating the Patient menu Using the Vpap Adapt SVLearn Circuit Settings MenuMask Leak Alert Ramp Cpap mode onlyResults Menu SmartStopOptions Menu Servicing Menu Treatment ScreensPage Triggered by AlarmsLow Mask Pressure Alarm If a low pressure alarm soundsHigh Leak Alarm Power Fail AlarmHigh Pressure Alarm If a flow blocked alarm sounds Insufficient Pressure Support AlarmFlow Blocked Alarm If an insufficient pressure support alarm soundsPage Weekly Cleaning and MaintenanceDaily Periodically Air tubing assembly removed for cleaningFitting the filter Air FilterServicing Hypoallergenic Air FilterPage Unusual manner TroubleshootingPage Glossary AlarmsPage Technical Specifications IEC 601-1 classifications Inspiratory Trigger CharacteristicsPage Glossary of symbols Mask setting flow/pressure characteristicsEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index