ResMed Adapt SV System manual Humidifiers, Connecting a Humidifier

Page 17

Humidifiers

A humidifier may be required if you are experiencing dryness of the nose, throat,

or mouth. The following ResMed humidifiers are compatible with the VPAP Adapt SV:

HumidAire™

HumidAire 2i™

heated humidifier

heated humidifier

Humidifier accessory (HumidAire only)

Medium air tubing 21" (52 cm)

Figure 7: Humidifiers compatible with the VPAP Adapt SV

When using a humidifier, stop the VPAP Adapt SV by using the Start/Stop key or

by removing the mask to trigger the SmartStop function. Do not switch the VPAP Adapt SV off at the power switch or the power source while it is running, as

this could result in condensation in the motor.

Please refer to the relevant humidifier user manual for detailed instructions and warnings.

CAUTIONS

! Be very careful not to allow water to enter the VPAP Adapt SV as this could affect therapy and/or damage the device.

Do not tip the humidifier while it is connected to the VPAP Adapt SV. If water enters the VPAP Adapt SV device, turn off the main power switch at the back of the unit and unplug it from the power outlet. Return the unit to your equipment supplier for checking.

Connecting a Humidifier

Note: After connecting or removing a humidifier, always perform the Air Circuit Learn function (see page 22).

WARNING

! When using a humidifier, position it so that it is lower than your head will be during sleep, and at the same level or lower than the VPAP Adapt SV.

VPAP Adapt SV System

11

Image 17
Contents Vpap Adapt SV USA ENG Reorder number 26869/2 06 Page Contents Daily Weekly Periodically Air Filter Vpap Adapt SV IntroductionUser/Owner Responsibility DefinitionsPage Page Page Quick Setup Guide Page Vpap Adapt SV Components Vpap Adapt SV System Air tubing fully assembled MasksAssembling the Vpap Adapt SV System Connect the power cordAttach the mask to the air tubing Humidifiers Connecting a HumidifierHumidAire 2i attached to the Vpap Adapt SV HumidAireHumidAire Page Anti-bacterial Filter ResLinkSupplemental Oxygen Using DC Power to Run the Vpap Adapt SVStandby Mode Using the Vpap Adapt SVStarting Treatment Stopping TreatmentLCD screen Control Panel FunctionsStart/Stop key Up/Down keyAlarm LEDs and Alarm Mute Adjusting SettingsLeft key green Right key redExample You wish to change the setting for the mask you are usingMenus Navigating the Patient menu Using the Vpap Adapt SVLearn Circuit Settings MenuMask Results Menu Ramp Cpap mode onlySmartStop Leak AlertOptions Menu Servicing Menu Treatment ScreensPage Low Mask Pressure Alarm AlarmsIf a low pressure alarm sounds Triggered byHigh Leak Alarm Power Fail AlarmHigh Pressure Alarm Flow Blocked Alarm Insufficient Pressure Support AlarmIf an insufficient pressure support alarm sounds If a flow blocked alarm soundsPage Weekly Cleaning and MaintenanceDaily Periodically Air tubing assembly removed for cleaningServicing Air FilterHypoallergenic Air Filter Fitting the filterPage Unusual manner TroubleshootingPage Glossary AlarmsPage Technical Specifications IEC 601-1 classifications Inspiratory Trigger CharacteristicsPage Glossary of symbols Mask setting flow/pressure characteristicsEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index