ResMed Adapt SV System manual Emissions test Compliance

Page 49

Guidance and Manufacturer’s Declaration — Electromagnetic Emissions and Immunity

Guidance and manufacturer’s declaration – electromagnetic emissions

The VPAP Adapt SV is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP Adapt SV should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment -

guidance

 

 

 

 

 

RF emissions CISPR11

Group 1

The VPAP Adapt SV uses RF energy

 

 

only for its internal function.

 

 

Therefore, its RF emissions are very

 

 

low and are not likely to cause any

 

 

interference in nearby electronic

 

 

equipment.

 

 

 

RF emissions CISPR 11

Class B

The VPAP Adapt SV is suitable for use

 

 

in all establishments, including

 

 

Harmonic Emissions

Class A

domestic establishments and those

IEC 61000-3-2

 

directly connected to the public low-

 

 

voltage network that supplies

 

 

Voltage Fluctuations/Flicker

Complies

buildings used for domestic purposes.

Emissions IEC 61000-3-3

 

 

 

 

 

Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to EMC information provided in this document.

Warnings: The VPAP Adapt SV should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the VPAP Adapt SV should be observed to verify normal operation in the configuration in which it will be used.

The use of accessories (eg Humidifiers) other than those specified in this manual is not recommended. They may result in increased emissions or immunity of the VPAP Adapt SV.

Technical Specifications

43

Image 49
Contents Vpap Adapt SV USA ENG Reorder number 26869/2 06 Page Contents Daily Weekly Periodically Air Filter Vpap Adapt SV IntroductionUser/Owner Responsibility DefinitionsPage Page Page Quick Setup Guide Page Vpap Adapt SV Components Vpap Adapt SV SystemAir tubing fully assembled MasksAssembling the Vpap Adapt SV System Connect the power cordAttach the mask to the air tubing Humidifiers Connecting a HumidifierHumidAire 2i attached to the Vpap Adapt SV HumidAireHumidAire Page Anti-bacterial Filter ResLinkSupplemental Oxygen Using DC Power to Run the Vpap Adapt SVStandby Mode Using the Vpap Adapt SVStarting Treatment Stopping TreatmentLCD screen Control Panel FunctionsStart/Stop key Up/Down keyAlarm LEDs and Alarm Mute Adjusting SettingsLeft key green Right key redExample You wish to change the setting for the mask you are usingMenus Navigating the Patient menu Using the Vpap Adapt SVMask Settings MenuLearn Circuit Results Menu Ramp Cpap mode onlySmartStop Leak AlertOptions Menu Servicing Menu Treatment ScreensPage Low Mask Pressure Alarm AlarmsIf a low pressure alarm sounds Triggered byHigh Pressure Alarm Power Fail AlarmHigh Leak Alarm Flow Blocked Alarm Insufficient Pressure Support AlarmIf an insufficient pressure support alarm sounds If a flow blocked alarm soundsPage Daily Cleaning and MaintenanceWeekly Periodically Air tubing assembly removed for cleaningServicing Air FilterHypoallergenic Air Filter Fitting the filterPage Unusual manner TroubleshootingPage Glossary AlarmsPage Technical Specifications IEC 601-1 classifications Inspiratory Trigger CharacteristicsPage Glossary of symbols Mask setting flow/pressure characteristicsEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index