ResMed Adapt SV System manual Power Fail Alarm, High Pressure Alarm, High Leak Alarm

Page 34

Will clear when:

the low pressure condition is fixed

treatment is stopped by pressing the Start/Stop key.

Power Fail Alarm

Triggered by:

power failure

machine is disconnected or switched off while delivering treatment.

If a power fail alarm sounds:

the flow generator stops delivering air pressure.

Will stop:

when the Alarm Mute key is pressed, or

after two minutes, or

when power is restored.

CAUTION

! Remove the mask from your face if the power fails.

High Pressure Alarm

Triggered by:

the mask pressure exceeds the set trigger level (25 cm H2O) for more than 700 ms.

If a high pressure alarm sounds:

1.The treatment will stop.

2.Turn power off.

3.Check that the air tube and sensor tubes are connected properly.

4.Turn power back on.

5.Remove mask and perform Air Circuit Learn function.

6.Try using the flow generator one more time.

7.If the high pressure alarm activates repeatedly, discontinue use and return to ResMed for servicing. If the alarm does not recur, then continue to use as normal.

Note: The alarm system is designed to ignore coughing. However, if a cough is especially intense and prolonged, it may trigger the high pressure alarm.

Will stop when:

the VPAP Adapt SV is turned off.

High Leak Alarm

Triggered by:

high mask leak (greater than 30 L/min) for more than 20 seconds

the mask being removed while SmartStop is disabled.

If a high leak alarm sounds:

adjust the mask to minimize leak

28

Image 34
Contents Vpap Adapt SV USA ENG Reorder number 26869/2 06 Page Contents Daily Weekly Periodically Air Filter User/Owner Responsibility IntroductionVpap Adapt SV DefinitionsPage Page Page Quick Setup Guide Page Vpap Adapt SV System Vpap Adapt SV ComponentsMasks Air tubing fully assembledConnect the power cord Assembling the Vpap Adapt SV SystemAttach the mask to the air tubing Connecting a Humidifier HumidifiersHumidAire HumidAire 2i attached to the Vpap Adapt SVHumidAire Page ResLink Anti-bacterial FilterUsing DC Power to Run the Vpap Adapt SV Supplemental OxygenStarting Treatment Using the Vpap Adapt SVStandby Mode Stopping TreatmentStart/Stop key Control Panel FunctionsLCD screen Up/Down keyLeft key green Adjusting SettingsAlarm LEDs and Alarm Mute Right key redYou wish to change the setting for the mask you are using ExampleNavigating the Patient menu Using the Vpap Adapt SV MenusMask Settings MenuLearn Circuit SmartStop Ramp Cpap mode onlyResults Menu Leak AlertOptions Menu Treatment Screens Servicing MenuPage If a low pressure alarm sounds AlarmsLow Mask Pressure Alarm Triggered byHigh Pressure Alarm Power Fail AlarmHigh Leak Alarm If an insufficient pressure support alarm sounds Insufficient Pressure Support AlarmFlow Blocked Alarm If a flow blocked alarm soundsPage Daily Cleaning and MaintenanceWeekly Air tubing assembly removed for cleaning PeriodicallyHypoallergenic Air Filter Air FilterServicing Fitting the filterPage Troubleshooting Unusual mannerPage Alarms GlossaryPage Technical Specifications Inspiratory Trigger Characteristics IEC 601-1 classificationsPage Mask setting flow/pressure characteristics Glossary of symbolsEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index