Guidance and manufacturer’s declaration – electromagnetic immunity
The VPAP Adapt SV is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP Adapt SV should assure that it is used in such an environment.
| Compliance |
| Electromagnetic | |
Immunity test |
| environment – | ||
test level | level |
| ||
|
| guidance | ||
|
|
|
| |
|
|
|
|
|
Electrostatic | ±6 kV contact | ±6 kV contact |
| Floors should be wood, |
discharge (ESD) |
|
|
| concrete or ceramic tile. If |
IEC | ±8 kV air | ±8 kV air |
| floors are covered with |
|
|
|
| synthetic material, the |
|
|
|
| relative humidity should be |
|
|
|
| at least 30%. |
|
|
|
|
|
Electrical fast | ±2 kV for power | ±2 kV |
| Mains power quality should |
transient/burst | supply lines |
|
| be that of a typical |
IEC |
|
|
| commercial or hospital |
| ±1 kV for input/ | Not Applicable |
| environment. |
| output lines |
|
|
|
|
|
|
|
|
Surge | ±1 kV differential | ±1 kV differential |
| Mains power quality should |
IEC | mode | mode |
| be that of a typical |
|
|
|
| commercial or hospital |
| ±2 kV common | ±2 kV common |
| environment. |
| mode | mode |
|
|
|
|
|
|
|
Voltage dips, short | <5% Ut | < 12 V |
| Mains power quality should |
interruptions and | (>95% dip in Ut) | (>95% dip in 240V) |
| be that of a typical |
voltage variations on | for 0.5 cycle | for 0.5 cycle |
| commercial or hospital |
power supply input |
|
|
| environment. |
lines. | 40% Ut | 96 V |
| If the user of the |
IEC | (60% dip in Ut) | (60% dip in 240V) |
| VPAP Adapt SV requires |
| for 5 cycles | for 5 cycles |
| continued operation during |
|
|
|
| power mains interruptions, |
| 70% Ut | 168 V |
| it is recommended that the |
| (30% dip in Ut) | (30% dip in 240V) |
| VPAP Adapt SV be powered |
| for 25 cycles | for 25 cycles |
| from an uninterruptible |
| <5% Ut | <12 V |
| power source |
|
|
| ||
| (>95%dip in Ut) | (>95%dip in 240V) |
|
|
| for 5 sec | for 5 sec |
|
|
|
|
|
|
|
Power frequency | 3 A/m | 3 A/m |
| Power frequency magnetic |
(50/60 Hz) |
|
|
| fields should be at levels |
magnetic field |
|
|
| characteristic of a typical |
IEC |
|
|
| location in a typical |
|
|
|
| commercial or hospital |
|
|
|
| environment |
|
|
|
|
|
NOTE: Ut is the AC mains voltage prior to application of the test level. |
| |||
|
|
|
|
|
(Continued next page)