ResMed Adapt SV System manual Portable and mobile RF communications

Page 51

Guidance and manufacturer’s declaration – electromagnetic immunity (continued)

The VPAP Adapt SV is intended for use in the electromagnetic environment specified below. The customer or the user of the VPAP Adapt SV should assure that it is used in such an environment.

Immunity

IEC60601-1-2

Complianc

 

Electromagnetic environment –

test

test level

e level

 

guidance

 

 

 

 

 

 

 

 

 

 

Portable and mobile RF communications

 

 

 

 

equipment should be used no closer to any

 

 

 

 

part of the VPAP Adapt SV, including cables,

 

 

 

 

than the recommended separation distance

 

 

 

 

calculated from the equation applicable to the

 

 

 

 

frequency of the transmitter.

 

 

 

 

Recommended separation distance

Conducted RF

3 Vrms

3 Vrms

 

d = 1.17 P

IEC 61000-4-6

150 kHz to 80 MHz

 

 

 

 

Radiated RF

10 V/m

10 V/m

 

d = 0.35 P 80 MHz to 800 MHz

IEC 61000-4-3

80 MHz to 2.5 GHz

 

 

 

 

 

 

 

 

d = 0.70 P 800 MHz to 2.5 GHz

 

 

 

 

where P is the maximum output power rating

 

 

 

 

of the transmitter in watts (W) according to the

 

 

 

 

transmitter manufacturer and d is the

 

 

 

 

recommended separation distance in meters

 

 

 

 

(m).

 

 

 

 

Field strengths from fixed RF transmitters, as

 

 

 

 

determined by an electromagnetic site

 

 

 

 

survey,a should be less than the compliance

 

 

 

 

level in each frequency range.b

 

 

 

 

Interference may occur in the vicinity of

 

 

 

 

equipment marked with the following symbol:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the VPAP Adapt SV is used exceeds the applicable RF compliance level above, the VPAP Adapt SV should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the VPAP Adapt SV.

bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

Technical Specifications

45

Image 51
Contents Vpap Adapt SV USA ENG Reorder number 26869/2 06 Page Contents Daily Weekly Periodically Air Filter Definitions IntroductionVpap Adapt SV User/Owner ResponsibilityPage Page Page Quick Setup Guide Page Vpap Adapt SV Components Vpap Adapt SV SystemAir tubing fully assembled MasksAssembling the Vpap Adapt SV System Connect the power cordAttach the mask to the air tubing Humidifiers Connecting a HumidifierHumidAire 2i attached to the Vpap Adapt SV HumidAireHumidAire Page Anti-bacterial Filter ResLinkSupplemental Oxygen Using DC Power to Run the Vpap Adapt SVStopping Treatment Using the Vpap Adapt SVStandby Mode Starting TreatmentUp/Down key Control Panel FunctionsLCD screen Start/Stop keyRight key red Adjusting SettingsAlarm LEDs and Alarm Mute Left key greenExample You wish to change the setting for the mask you are usingMenus Navigating the Patient menu Using the Vpap Adapt SVSettings Menu MaskLearn Circuit Leak Alert Ramp Cpap mode onlyResults Menu SmartStopOptions Menu Servicing Menu Treatment ScreensPage Triggered by AlarmsLow Mask Pressure Alarm If a low pressure alarm soundsPower Fail Alarm High Pressure AlarmHigh Leak Alarm If a flow blocked alarm sounds Insufficient Pressure Support AlarmFlow Blocked Alarm If an insufficient pressure support alarm soundsPage Cleaning and Maintenance DailyWeekly Periodically Air tubing assembly removed for cleaningFitting the filter Air FilterServicing Hypoallergenic Air FilterPage Unusual manner TroubleshootingPage Glossary AlarmsPage Technical Specifications IEC 601-1 classifications Inspiratory Trigger CharacteristicsPage Glossary of symbols Mask setting flow/pressure characteristicsEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index