ResMed Adapt SV System manual HumidAire

Page 19

HumidAire™

You will need a medium size 21" (52 cm) air tube to connect the VPAP Adapt SV unit to the HumidAire™ (see Figure 7).

1Make sure both the HumidAire and the VPAP Adapt SV are turned off.

Fill the HumidAire with water as described in the humidifier manual. Place the filled water chamber inside the HumidAire.

2Gently detach enough of the tube clips to release the end of the pressure sensor tube from the VPAP Adapt SV air tubing.

Connect the VPAP Adapt SV air tubing to the left connector port on the humidifier,

and the medium 21" (52 cm) air tubing to the right connector port. Close the

HumidAire lid.

Medium air tubing

HumidAire

VPAP Adapt SV air tubing

Pressure sensor tube

3Place the VPAP Adapt SV on top of the HumidAire. Never place the

VPAP Adapt SV unit underneath the humidifier. (This is to avoid water spilling into the unit.)

4Connect the free end of the medium air tubing to the VPAP Adapt SV.

Connect the pressure sensor tube by twisting the Luer lock onto the VPAP Adapt SV (see page 10).

VPAP Adapt SV System

13

Image 19
Contents Vpap Adapt SV USA ENG Reorder number 26869/2 06 Page Contents Daily Weekly Periodically Air Filter Definitions IntroductionVpap Adapt SV User/Owner ResponsibilityPage Page Page Quick Setup Guide Page Vpap Adapt SV Components Vpap Adapt SV SystemAir tubing fully assembled MasksAssembling the Vpap Adapt SV System Connect the power cordAttach the mask to the air tubing Humidifiers Connecting a HumidifierHumidAire 2i attached to the Vpap Adapt SV HumidAireHumidAire Page Anti-bacterial Filter ResLinkSupplemental Oxygen Using DC Power to Run the Vpap Adapt SVStopping Treatment Using the Vpap Adapt SVStandby Mode Starting TreatmentUp/Down key Control Panel FunctionsLCD screen Start/Stop keyRight key red Adjusting SettingsAlarm LEDs and Alarm Mute Left key greenExample You wish to change the setting for the mask you are usingMenus Navigating the Patient menu Using the Vpap Adapt SVMask Settings MenuLearn Circuit Leak Alert Ramp Cpap mode onlyResults Menu SmartStopOptions Menu Servicing Menu Treatment ScreensPage Triggered by AlarmsLow Mask Pressure Alarm If a low pressure alarm soundsHigh Pressure Alarm Power Fail AlarmHigh Leak Alarm If a flow blocked alarm sounds Insufficient Pressure Support AlarmFlow Blocked Alarm If an insufficient pressure support alarm soundsPage Daily Cleaning and MaintenanceWeekly Periodically Air tubing assembly removed for cleaningFitting the filter Air FilterServicing Hypoallergenic Air FilterPage Unusual manner TroubleshootingPage Glossary AlarmsPage Technical Specifications IEC 601-1 classifications Inspiratory Trigger CharacteristicsPage Glossary of symbols Mask setting flow/pressure characteristicsEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index