Emerson Process Management 2051 manual Sensor Transducer Block, Rosemount, Reference Manual

Page 66
SENSOR TRANSDUCER BLOCK

Rosemount 2051

Reference Manual

00809-0200-4101, Rev AA July 2008

SENSOR TRANSDUCER BLOCK

Table 5-5. Sensor Transducer Block BLOCK_ERR messages

Table 5-6. Sensor Transducer Block XD_ERR messages

This section describes error conditions found in the Sensor Transducer Block. Read Table 5-5 through Table 5-7 to determine the appropriate corrective action.

Condition Name and Description

Other

Out of Service: The actual mode is out of service.

Condition Name and Description

Electronics Failure: An electrical component failed.

I/O Failure: An I/O failure occurred.

Data Integrity Error: Data stored in the device is no longer valid due to a non-volatile memory checksum failure, a data verify after write failure, etc.

Algorithm Error: The algorithm used in the transducer block produced an error due to overflow, data reasonableness failure, etc.

Diagnostics

Table 5-7 lists the potential errors and the possible corrective actions for the given values. The corrective actions are in order of increasing system level compromises. The first step should always be to reset the transmitter and then if the error persists, try the steps in Table 5-7. Start with the first corrective action and then try the second.

Table 5-7. Sensor Transducer Block DETAILED_XD_STATUS and RECOMMENDED_ACTION messages

Condition Name and Description

 

RECOMMENDED_ACTION

 

 

Pressure sensor not updating

1. Restart Processor

 

2.

Reconnect sensor module cable

 

3.

Send to Service Center

 

 

Temperature sensor not updating

1. Restart Processor

 

2.

Reconnect sensor module cable

 

3.

Send to Service Center

Sensor ROM Check sum failure

1. Restart Processor

 

2.

Send to Service Center

 

 

Sensor NV write failure

1. Restart Processor

 

2.

Send to Service Center

Sensor RAM check sum error

1. Restart Processor

 

2.

Send to Service Center

 

 

Sensor NV factory data warning

1. Restart Processor

 

2.

Send to Service Center

Sensor NV user data warning

1. Restart Processor

 

2.

Send to Service Center

 

 

Sensor NV user data error

1. Restart Processor

 

2.

Send to Service Center

Sensor NV factory data error

1. Restart Processor

 

2.

Send to Service Center

 

 

Pressure sensor out of limits

1. Check pressure

 

2.

Restart Processor

Sensor temperature out of limits

1. Check Temperature

 

2.

Restart Processor

 

 

Sensor temperature beyond failure limits

1. Check Temperature

 

2.

Restart Processor

 

3.

Send to Service Center

 

 

 

5-6

Image 66
Contents with FOUNDATION Fieldbus Protocol Reference ManualRosemount 2051 Pressure Transmitter Page Rosemount Rosemount 2051 PressureTransmitter with FOUNDATION Fieldbus Reference ManualPage SECTION 3 Configuration Table of ContentsSECTION 1 Introduction SECTION 2 Installation RosemountAPPENDIX A Reference Data SECTION Operation and MaintenanceSECTION 5 Troubleshooting RosemountRosemount APPENDIX B Approval InformationAPPENDIX C Block Information Reference ManualRosemount Reference ManualTOC-4 Introduction USING THIS MANUAL SERVICE SUPPORTSection RosemountMODELS COVERED RosemountReference Manual Section InstallationOVERVIEW SAFETY MESSAGES Warnings RosemountMECHANICAL GENERALCONSIDERATIONS ENVIRONMENTAL2051C Coplanar Flange Dimensional Drawing INSTALLATION PROCEDURESDimensional Drawings RosemountReference Manual Coplanar Flange Mounting Configurations withRosemount 00809-0200-4101,Rev AA2051C Coplanar with Traditional Flange RosemountReference Manual Rosemount Traditional Flange Mounting Configurations withOptional Brackets for 2-in.Pipe or Panel Mounting Reference ManualRosemount 2051T Dimensional Drawings2051T with Rosemount 306 Integral Manifold Reference ManualPipe Mounting RosemountReference Manual Panel Mounting2051L Liquid Level RosemountReference Manual Mount the Transmitter RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Reference Manual Figure 2-6.Traditional Flange Bolt ConfigurationsRosemount 00809-0200-4101,Rev AA JulyImpulse Piping RosemountReference Manual Rosemount Reference ManualFlow Process Connections RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Housing Rotation RosemountReference Manual Rosemount LCD DisplayTagging Reference ManualRosemount Conduit InstallationELECTRICAL Reference ManualWiring RosemountReference Manual Transient Protection Terminal Block RosemountReference Manual JumpersSecurity RosemountReference Manual Grounding RosemountReference Manual Rosemount Figure 2-17.External Ground AssemblyExternal Ground Assembly Reference ManualRosemount HAZARDOUS LOCATIONS CERTIFICATIONSROSEMOUNT 305, 306 AND 304 MANIFOLDS Reference ManualManifold Installation Procedure Rosemount Rosemount 305 Integral Manifold InstallationProcedure Rosemount 306 Integral Conventional Manifold Installation ProcedureIntegral Manifold Operation RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount LIQUID LEVEL MEASUREMENTOpen Vessels Closed Vessels Reference ManualRosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference ManualSection ConfigurationOVERVIEW SAFETY MESSAGES Warnings RosemountGENERAL BLOCK INFORMATION Modes DEVICE CAPABILITIES Link Active SchedulerCapabilities RosemountRosemount RESOURCE BLOCK FEATURES and FEATURES SELSimulation Reference ManualMAX_NOTIFY RosemountReference Manual Rosemount Configure the AI blockANALOG INPUT AI FUNCTION BLOCK Reference ManualConfiguration Examples RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount FilteringLow Cutoff Reference ManualRosemount Process Alarms Alarm PriorityStatus Options Reference ManualAdvanced Features LCD BLOCK RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Section 4 Operation and MaintenanceOVERVIEW SAFETY MESSAGES Warnings Reference ManualSTATUS Master Reset Method Simulation RosemountReference Manual Rosemount CALIBRATION Sensor CalibrationUpper and Lower Trim Methods Reference ManualRosemount Sensor Calibration, Zero Trim MethodFactory Calibration Method Reference ManualRosemount Section 5 TroubleshootingOVERVIEW SAFETY MESSAGES Warnings Reference ManualReference Manual TROUBLESHOOTING GUIDESRosemount 00809-0200-4101,Rev AA JulyReference Manual Figure 5-2.Problems with communications flowchartRosemount 00809-0200-4101,Rev AA JulyReference Manual Table 5-1.Troubleshooting guideRosemount 00809-0200-4101,Rev AA JulyRESOURCE BLOCK RosemountReference Manual SENSOR TRANSDUCER BLOCK RosemountReference Manual Reference Manual LCD TRANSDUCER BLOCKRosemount ANALOG INPUT AI FUNCTION BLOCKReference Manual 1.Put Resource block into OOS Out of ServiceRosemount 00809-0200-4101,Rev AA JulyRosemount DISASSEMBLY PROCEDURES Remove from ServiceRemove Terminal Block Reference ManualRemove the Electronics Board RosemountReference Manual Reassemble the 2051C Process Flange REASSEMBLY PROCEDURES Attach theElectronics Board Install the Terminal Block RosemountInstall the Drain/Vent Valve RosemountReference Manual Appendix A PERFORMANCE SPECIFICATIONSConformance To Specification ±3σ Sigma Reference DataLine Pressure Effect per 1000 psi 6,9 MPa Long Term StabilityDynamic Performance RosemountMounting Position Effects Vibration Effect Power Supply EffectAmbient Temperature Effect per 50F 28C Electromagnetic Compatibility EMCProtocols FUNCTIONAL SPECIFICATIONS Range and Sensor LimitsService RosemountRosemount Reference Manual00809-0200-4101,Rev AA July Overpressure Limits RosemountReference Manual Rosemount Static Pressure Limit Burst Pressure LimitsTemperature Limits Reference ManualRosemount Humidity Limits Volumetric Displacement DampingFailure Mode Alarm Reference Manual2051T Process Wetted Parts PHYSICAL SPECIFICATIONS Electrical ConnectionsProcess Connections 2051C Process Wetted Parts 2051L Process Wetted PartsNon-WettedParts for 2051C/T/L RosemountReference Manual Table A-5.2051L Weights without Options Shipping WeightsTable A-4.Transmitter Weights without Options Table A-6.Transmitter Options WeightsORDERING INFORMATION RosemountReference Manual Rosemount Reference ManualAlternate Process Connection: Flange2 Rosemount Reference ManualCSA FISCO Intrinsically Safe Rosemount Reference ManualA-15 Rosemount Reference ManualGGage AAbsolute Rosemount Reference ManualFM Intrinsically Safe, Division Special Certifications Q4 Calibration Certificate RosemountReference Manual Prior-usecertificate of FMEDA dataRosemount Reference Manual4 -300to 300 psi/3 psi -20,7to 20,7 bar/0,2 bar Rosemount Reference ManualSyltherm XLT ATEX Flameproof, Intrinsic Safety, Type n, Dust RosemountReference Manual OPTIONS RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Reference Manual SPARE PARTSRosemount LCD Display, FOUNDATION fieldbusRosemount Reference ManualNickel-platedcarbon steel Rosemount Reference ManualGage Flange and Adapter Bolt Kit Set of Rosemount Reference ManualA-28 Warnings OVERVIEW SAFETY MESSAGESAppendix B Approval Information Approved Manufacturing LocationsHazardous Locations Certifications European Directive InformationFIELDBUS PROTOCOL RosemountRosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July APPROVAL DRAWINGS Factory Mutual FM RosemountReference Manual Rosemount Reference ManualRosemount Reference ManualB-10 Rosemount Reference ManualB-11 Rosemount Reference ManualB-12 Rosemount Reference ManualB-13 Rosemount Reference ManualB-14 Rosemount Reference ManualB-15 Rosemount Reference ManualB-16 Rosemount Reference ManualB-17 Rosemount Reference ManualB-18 Rosemount Reference ManualB-19 Rosemount Reference ManualB-20 Reference Manual Canadian Standards Association CSARosemount B-21Rosemount Reference ManualB-22 Rosemount Reference ManualB-23 Rosemount Reference ManualB-24 Rosemount Reference ManualB-25 Rosemount Reference ManualB-26 Rosemount Reference ManualB-27 Rosemount Reference ManualB-28 Rosemount Reference ManualB-29 Rosemount Reference ManualB-30 TRANSDUCER BLOCK Overview Appendix CBlock Information RosemountParameters and Descriptions RosemountReference Manual Rosemount Reference ManualPRIMARY_VALUE_RANGE Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount TroubleshootingOverview Reference ManualParameters and Descriptions RosemountReference Manual 00809-0200-4101,Rev AA JulyRosemount Reference ManualFLANGE_MTL Rosemount Reference ManualREM_SEAL_FILL Rosemount Reference ManualSimulate_STATE Reference Manual Block ErrorsRosemount 00809-0200-4101,Rev AA JulyRosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference ManualMode will not leave Rosemount Reference ManualC-14 Index RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference ManualIndex-4 Rosemount Reference Manual00809-0200-4101,Rev AA July NOTES Rosemount Reference ManualPage Rosemount ¢00809-0100-4801V¤Reference Manual