Install the Drain/Vent Valve

Rosemount 2051

Reference Manual

00809-0200-4101, Rev AA July 2008

a. Coplanar Process Flange:

Hold the process flange in place by installing the two alignment screws to finger tightness (screws are not pressure retaining). Do not overtighten as this will affect module-to-flange alignment.

Install the four 1.75-in. flange bolts by finger tightening them to the flange.

Table 5-11. Bolt Installation Torque Values

b. Coplanar Process Flange with Flange Adapters:

Hold the process flange in place by installing the two alignment screws to finger tightness (screws are not pressure retaining). Do not overtighten as this will affect module-to-flange alignment.

Hold the flange adapters and adapter o-rings in place while installing the four configurations, use four 2.88-in. bolts. For gage pressure configurations, use two 2.88-in. bolts and two 1.75-in. bolts.

c. Manifold:

Contact the manifold manufacturer for the appropriate bolts and procedures.

3.Tighten the bolts to the initial torque value using a crossed pattern. See Table 5-11 for appropriate torque values.

Bolt Material

Initial Torque Value

Final Torque Value

 

 

 

CS-ASTM-A449 Standard

300 in-lb. (34 N-m)

650 in-lb. (73 N-m)

 

 

 

316 SST—Option L4

150 in-lb. (17 N-m)

300 in-lb. (34 N-m)

 

 

 

ASTM-A-193-B7M—Option

300 in-lb. (34 N-m)

650 in-lb. (73 N-m)

L5

 

 

 

 

 

ASTM-A-193 Class 2, Grade

150 in.-lb (17 N-m)

300 in.-lb (34 N-m)

B8M—Option L8

 

 

 

 

 

Install the Drain/Vent Valve

NOTE

If you replaced the PTFE sensor module o-rings, re-torque the flange bolts after installation to compensate for cold flow.

NOTE

After replacing o-rings on Range 1 transmitters and re-installing the process flange, expose the transmitter to a temperature of 185 °F (85 °C) for two hours. Then re-tighten the flange bolts in a cross pattern, and again expose the transmitter to a temperature of 185 °F (85 °C) for two hours before calibration.

1.Apply sealing tape to the threads on the seat. Starting at the base of the valve with the threaded end pointing toward the installer, apply two clockwise turns of sealing tape.

2.Tighten the drain/vent valve to 250 in-lb. (28.25 N-m).

3.Take care to place the opening on the valve so that process fluid will drain toward the ground and away from human contact when the valve is opened.

5-12

Image 72
Contents with FOUNDATION Fieldbus Protocol Reference ManualRosemount 2051 Pressure Transmitter Page Rosemount 2051 Pressure Transmitter with FOUNDATION FieldbusRosemount Reference ManualPage Table of Contents SECTION 1 Introduction SECTION 2 InstallationSECTION 3 Configuration RosemountSECTION Operation and Maintenance SECTION 5 TroubleshootingAPPENDIX A Reference Data RosemountAPPENDIX B Approval Information APPENDIX C Block InformationRosemount Reference ManualRosemount Reference ManualTOC-4 USING THIS MANUAL SERVICE SUPPORT SectionIntroduction RosemountMODELS COVERED RosemountReference Manual Installation OVERVIEW SAFETY MESSAGES WarningsSection RosemountGENERAL CONSIDERATIONSMECHANICAL ENVIRONMENTALINSTALLATION PROCEDURES Dimensional Drawings2051C Coplanar Flange Dimensional Drawing RosemountCoplanar Flange Mounting Configurations with RosemountReference Manual 00809-0200-4101,Rev AA2051C Coplanar with Traditional Flange RosemountReference Manual Traditional Flange Mounting Configurations with Optional Brackets for 2-in.Pipe or Panel MountingRosemount Reference Manual2051T Dimensional Drawings 2051T with Rosemount 306 Integral ManifoldRosemount Reference ManualRosemount Reference ManualPipe Mounting Panel Mounting2051L Liquid Level RosemountReference Manual Mount the Transmitter RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Figure 2-6.Traditional Flange Bolt Configurations RosemountReference Manual 00809-0200-4101,Rev AA JulyImpulse Piping RosemountReference Manual Rosemount Reference ManualFlow Process Connections RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Housing Rotation RosemountReference Manual LCD Display TaggingRosemount Reference ManualConduit Installation ELECTRICALRosemount Reference ManualWiring RosemountReference Manual Transient Protection Terminal Block RosemountReference Manual JumpersSecurity RosemountReference Manual Grounding RosemountReference Manual Figure 2-17.External Ground Assembly External Ground AssemblyRosemount Reference ManualHAZARDOUS LOCATIONS CERTIFICATIONS ROSEMOUNT 305, 306 AND 304 MANIFOLDSRosemount Reference ManualRosemount 305 Integral Manifold Installation Procedure Rosemount 306 IntegralManifold Installation Procedure Rosemount Conventional Manifold Installation ProcedureIntegral Manifold Operation RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July LIQUID LEVEL MEASUREMENT Open Vessels Closed VesselsRosemount Reference ManualRosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference ManualConfiguration OVERVIEW SAFETY MESSAGES WarningsSection RosemountDEVICE CAPABILITIES Link Active Scheduler CapabilitiesGENERAL BLOCK INFORMATION Modes RosemountRESOURCE BLOCK FEATURES and FEATURES SEL SimulationRosemount Reference ManualMAX_NOTIFY RosemountReference Manual Configure the AI block ANALOG INPUT AI FUNCTION BLOCKRosemount Reference ManualConfiguration Examples RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Filtering Low CutoffRosemount Reference ManualProcess Alarms Alarm Priority Status OptionsRosemount Reference ManualAdvanced Features LCD BLOCK RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Section 4 Operation and Maintenance OVERVIEW SAFETY MESSAGES WarningsRosemount Reference ManualSTATUS Master Reset Method Simulation RosemountReference Manual CALIBRATION Sensor Calibration Upper and Lower Trim MethodsRosemount Reference ManualSensor Calibration, Zero Trim Method Factory Calibration MethodRosemount Reference ManualSection 5 Troubleshooting OVERVIEW SAFETY MESSAGES WarningsRosemount Reference ManualTROUBLESHOOTING GUIDES RosemountReference Manual 00809-0200-4101,Rev AA JulyFigure 5-2.Problems with communications flowchart RosemountReference Manual 00809-0200-4101,Rev AA JulyTable 5-1.Troubleshooting guide RosemountReference Manual 00809-0200-4101,Rev AA JulyRESOURCE BLOCK RosemountReference Manual SENSOR TRANSDUCER BLOCK RosemountReference Manual LCD TRANSDUCER BLOCK RosemountReference Manual ANALOG INPUT AI FUNCTION BLOCK1.Put Resource block into OOS Out of Service RosemountReference Manual 00809-0200-4101,Rev AA JulyDISASSEMBLY PROCEDURES Remove from Service Remove Terminal BlockRosemount Reference ManualRemove the Electronics Board RosemountReference Manual REASSEMBLY PROCEDURES Attach the Electronics Board Install the Terminal BlockReassemble the 2051C Process Flange RosemountInstall the Drain/Vent Valve RosemountReference Manual PERFORMANCE SPECIFICATIONS Conformance To Specification ±3σ SigmaAppendix A Reference DataLong Term Stability Dynamic PerformanceLine Pressure Effect per 1000 psi 6,9 MPa RosemountVibration Effect Power Supply Effect Ambient Temperature Effect per 50F 28CMounting Position Effects Electromagnetic Compatibility EMCFUNCTIONAL SPECIFICATIONS Range and Sensor Limits ServiceProtocols RosemountRosemount Reference Manual00809-0200-4101,Rev AA July Overpressure Limits RosemountReference Manual Static Pressure Limit Burst Pressure Limits Temperature LimitsRosemount Reference ManualHumidity Limits Volumetric Displacement Damping Failure Mode AlarmRosemount Reference ManualPHYSICAL SPECIFICATIONS Electrical Connections Process Connections 2051C Process Wetted Parts2051T Process Wetted Parts 2051L Process Wetted PartsNon-WettedParts for 2051C/T/L RosemountReference Manual Shipping Weights Table A-4.Transmitter Weights without OptionsTable A-5.2051L Weights without Options Table A-6.Transmitter Options WeightsORDERING INFORMATION RosemountReference Manual Rosemount Reference ManualAlternate Process Connection: Flange2 Rosemount Reference ManualCSA FISCO Intrinsically Safe Rosemount Reference ManualA-15 Rosemount Reference ManualGGage AAbsolute Rosemount Reference ManualFM Intrinsically Safe, Division Rosemount Reference ManualSpecial Certifications Q4 Calibration Certificate Prior-usecertificate of FMEDA dataRosemount Reference Manual4 -300to 300 psi/3 psi -20,7to 20,7 bar/0,2 bar Rosemount Reference ManualSyltherm XLT ATEX Flameproof, Intrinsic Safety, Type n, Dust RosemountReference Manual OPTIONS RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July SPARE PARTS RosemountReference Manual LCD Display, FOUNDATION fieldbusRosemount Reference ManualNickel-platedcarbon steel Rosemount Reference ManualGage Flange and Adapter Bolt Kit Set of Rosemount Reference ManualA-28 OVERVIEW SAFETY MESSAGES Appendix B Approval InformationWarnings Approved Manufacturing LocationsEuropean Directive Information FIELDBUS PROTOCOLHazardous Locations Certifications RosemountRosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July APPROVAL DRAWINGS Factory Mutual FM RosemountReference Manual Rosemount Reference ManualRosemount Reference ManualB-10 Rosemount Reference ManualB-11 Rosemount Reference ManualB-12 Rosemount Reference ManualB-13 Rosemount Reference ManualB-14 Rosemount Reference ManualB-15 Rosemount Reference ManualB-16 Rosemount Reference ManualB-17 Rosemount Reference ManualB-18 Rosemount Reference ManualB-19 Rosemount Reference ManualB-20 Canadian Standards Association CSA RosemountReference Manual B-21Rosemount Reference ManualB-22 Rosemount Reference ManualB-23 Rosemount Reference ManualB-24 Rosemount Reference ManualB-25 Rosemount Reference ManualB-26 Rosemount Reference ManualB-27 Rosemount Reference ManualB-28 Rosemount Reference ManualB-29 Rosemount Reference ManualB-30 Appendix C Block InformationTRANSDUCER BLOCK Overview RosemountParameters and Descriptions RosemountReference Manual Rosemount Reference ManualPRIMARY_VALUE_RANGE Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Troubleshooting OverviewRosemount Reference ManualRosemount Reference ManualParameters and Descriptions 00809-0200-4101,Rev AA JulyRosemount Reference ManualFLANGE_MTL Rosemount Reference ManualREM_SEAL_FILL Rosemount Reference ManualSimulate_STATE Block Errors RosemountReference Manual 00809-0200-4101,Rev AA JulyRosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference ManualMode will not leave Rosemount Reference ManualC-14 Index RosemountReference Manual Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference Manual00809-0200-4101,Rev AA July Rosemount Reference ManualIndex-4 Rosemount Reference Manual00809-0200-4101,Rev AA July NOTES Rosemount Reference ManualPage Rosemount ¢00809-0100-4801V¤Reference Manual