Philips M5066A owner manual When emergency medical services arrive, Do not Delay Treatment

Page 24

3-6

If the patient is under 55 pounds or 8 years old, but you do NOT have an Infant/Child

Pads Cartridge:

DO NOT DELAY TREATMENT.

Remove all clothing from the torso, to bare both the chest and the back.

Apply the HeartStart using the adult pads cartridge, but place one pad in the center of the chest between the nipples, and the other in the center of the back (anterior-posterior).

If the patient is over 55 pounds or 8 years old, or if you are not sure of the exact weight or age:

DO NOT DELAY TREATMENT.

Remove all clothing from the chest.

Apply the HeartStart using the adult pads cartridge, and place the pads as illustrated on the pads (anterior-anterior). Make sure the pads do not overlap or touch each other.

When emergency medical services arrive

When Emergency Medical Services (EMS) personnel arrive to care for the patient, they may decide to apply another defibrillator to allow monitoring of the patient. The SMART Pads should be removed from the patient prior to using another defibrillator. EMS personnel may want a summary of the last-use data* stored in the HeartStart. To hear the summary data, hold down the i-button until the HeartStart beeps.

NOTE: After the EMS team removes the SMART Pads from the patient, remove the used pads cartridge, and insert a new pads cartridge before returning the HeartStart to service, to be sure it is ready for use.

* See Chapter 4, “After using the HeartStart” for details about data storage.

Philips Medical Systems

HEARTSTART M5066A

Image 24
Contents M5066A EditionIntentionally blank HeartStart Defibrillator M5066A Pads Cartridge HandleOn/Off Button. Press this Information Button. ThisIntentionally blank Quick Reference Intentionally blank HeartStart Important NoteIntentionally blank About this edition Edition historyCopyright Authorized EU representativeIntentionally blank Contents Appendices Introduction to the HeartStart DescriptionSudden Cardiac Arrest Indications for UseTraining and practice National and local requirementsFor more information Setting up the HeartStart Package contentsHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Using the HeartStart OverviewPull the green handle Place the pads Press the Shock button If a shock is neededTreating infants and children Remove the Infant/Child Pads Cartridge from its packageWhen emergency medical services arrive Do not Delay TreatmentPhilips Medical Systems Heartstart M5066A After using the HeartStart After each useHeartStart data storage Heartstart M5066A Routine Maintenance Maintaining the HeartStartPeriodic checks Cleaning the HeartStart Troubleshooting tipsDisposing of the HeartStart Accessories for the HeartStart REF 68-PCHATHeartstart M5066A Glossary of terms AEDCPR ECGNSA Smart NSA pause SCA Glossary of symbols/controls Symbol DescriptionDefibrillator and opens pads cartridge for use Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Precautions Contact with the patientRegularly for damage, according to directions Metal objects such as a bed frame or stretcherHeartStart Defibrillator specifications Technical informationPhysical EnvironmentalControls and indicators Category SealingDefibrillation Waveform Category Waveform parametersDelivered Current a Energy JDisarm AED mode Shock-to-shock cycle timeNewborn 14 J/kg ECG analysis system Category Function Shockable rhythms Non-shockable rhythmsAnalysis protocol ECG analysis performance Accessories specifications Battery M5070ACategory Specifications Battery type Prior to insertion Standby lifeEnvironmental considerations Whole and may also endanger human healthRegulations Configuration Parameter Speaker volumeDevice options ECG out dataPatient treatment protocol options NSA pause type Settings Standard NSADuring a Smart NSA pause, Pause HeartStart Monitoring duringCPR4 Installed Infants and children when an infantChild pads cartridge is installed Yes, NoHeartstart M5066A Testing and troubleshooting TestingTroubleshooting Troubleshooting while the HeartStart is in use Green Ready light is solid HeartStart tells youTo insert a pads cartridge HeartStart tells you To stop all motion Troubleshooting while the HeartStart is not in use Contact Philips for service if neededGreen Ready light is not on Behavior Electromagnetic conformity Electromagnetic emissionsEmissions test CisprElectromagnetic immunity Separation distance according to frequency of transmitter m Important warnings and reminders Environmental considerationsIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901